Introduction
Deficiencies are found in some specification
requirements for the use of standard flat bottom hole reference blocks
to perform periodic qualification calibrations of ultrasonic inspection
systems. Such performance checks often provide little or no indication
of the ability of a system to yield satisfactory results in actual material
inspections. A legitimate reason for the performance of periodic checks
is to discover signs of long term performance degradation that might
lead to costly delays during future inspections. A stable test block
assembly that has been used in such checking is described. Limitations
to the use of standard flat bottom hole reference blocks to check instrument
linearity or system sensitivity for material of other shapes, sizes
and composition is noted. Appropriate use of such blocks is for the
characterization of instrument performance without reference to any
specific inspection task.
The most fundamental, and perhaps only, measure
of the value of a nondestructive inspection system is the degree to
which its indications are representative of discontinuities that affect
the usefulness of the tested material for its intended purpose. The
phrase "representative of discontinuities" is the key part
of this sentence. It implies that the types, sizes, and orientations
of potential discontinuities that will affect material usefulness (that
is, quality) are known and that their seriousness may be estimated by
observing some properties of the indications produced by the examining
system. The properties most frequently measured for this purpose are
the amplitude, timing, and, sometimes, frequency content of the indications
as compared to the responses obtained from artificial targets in a reference
standard. Such standards (and the targets within them) are valid only
in so far as they represent the material to be inspected and actual
discontinuities that may be encountered in them. Ideally, the standard
is fabricated from an otherwise good sample of the material to be inspected.
Despite these limitations, there is merit to requiring a periodic
instrumentation check.
Many ultrasonic inspection specifications dictate two types of calibration
procedures to which an instrument or system must be subjected. These include
the usual preinspection calibration, in which the responses of the system
are measured to artificial targets in a reference standard.
The other type of calibration procedure often specified
is for a periodic qualifying check of an instrument or system. This
often involves measurement of the response to a set of reference block
standards. Although such standards, and the responses to them, may have
no correlation to the materials, number of discontinuities, or conditions
of any actual material inspection to be performed, this type of checking
procedure does have important uses as will be discussed below.
Preinspection Calibration and Checks During Inspection
The relationship of system indications to seriousness of the discontinuity
(size, location, etc.), that is, the sensitivity of the system, must,
of course, be estimated before inspection is begun. This is the purpose
of the preinspection calibration process.
Theoretical and experimental studies have been made
of the possibility of achieving accurate sensitivity threshold selection
for ultrasonic examinations without the need for conventional reference
standards (Gilmore and Czerw, 1977; Rogovsky and Rose, 1979 and 1980).
These methods use calculations based on the relationships among the
characteristics of transducers, instruments, typical discontinuities,
and materials. Although providing good insight into the interrelations
of these properties, this work has not led to widespread adoption of
the techniques described for actual inspections.
One way to calibrate the sensitivity of a system
for a given inspection might be to observe system responses while scanning
a large group of material samples that are somehow known to contain
discontinuities representative of various degrees of seriousness. Since
this is an obviously impractical, if not impossible, task to perform
prior to every inspection, an alternate procedure must be employed.
The most appropriate preinspection calibration consists
of observing system response to one or more artificially created anomalies,
or targets, of prescribed dimensions, locations, and orientations, in
an otherwise good sample of material of the same size, shape, surface
finish, and composition as the material to be inspected. Standard ASTM
type flat bottom hole reference blocks are sometimes specified to be
used as the standard. They are seldom appropriate for this function
since their shape, dimensions, and material properties are unlikely
to represent the actual material to be inspected and the types of discontinuities
to be detected.
This calibration procedure presupposes that the
artificial target characteristics have been selected from a database
somewhere that has established the validity of the rejection level(s)
so defined. The validation might have been established by correlating
(empirically or analytically) typical system responses to various artificial
discontinuities to those obtained from a group of known discontinuities.
The actual seriousness of the known discontinuities would preferably
have been itself established by subsequent destructive testing of the
samples observed.
In most cases, the characteristics of artificial
targets to serve as representative standards for system calibration
are selected on the basis of knowledge gained through many years of
successful screening of a given type of material for a given purpose
using systems calibrated by the response to particular types and sizes
of artificial targets.
Of course of equal importance to the selection of
artificial targets to represent minimum rejectable discontinuities is
having identical equipment and conditions during preinspection calibration
as are to be used for the actual testing. This obvious requirement is
a part of most test specifications.
Periodic System Qualification
The rationale for requiring a periodic instrument or system qualification
check is to ensure that the equipment has not suffered performance degradation
since its last check that might render it unable to detect the types
of discontinuities of interest.
System characteristics required to assure a desired
inspection sensitivity are verified automatically each time a preinspection
calibration, or calibration check during inspection, is made. This is
a much more realistic way of ensuring adequate performance than by depending
upon the results of periodic qualification checks using arbitrary reference
blocks.
An example of the type of qualification requirement
that does not relate significantly to system requirements for real discontinuity
detection is, for instance, the measurement of instrument linearity
by observing the amplitudes of signals from flat bottom hole (FBH) targets
of increasing area in a set of flat ASTM type reference blocks, even
though the system configuration and values that are necessary conditions
for many required inspections ignore well established criteria for the
legitimacy of such measurements (ASTM E 127, 1997; Beck, 1992):
- The entry surface of the material to be inspected
is curved while that of the blocks is flat.
- The size, frequency, and damping of the transducer
(as well as the bandwidth of the instrument) that are required to
produce an approximately linear area amplitude relationship, are not
appropriate to the type of inspections to be made.
- The actual accept/reject decisions to be made
do not require a linear amplitude relationship but involve only the
comparison of the amplitude of the signal from an indication to a
fixed threshold level determined by a reference target in the calibration
standard.
Other examples of typical system qualification requirements
that bear little or no relationship to the performance required for
discontinuity detection in actual test situations to be encountered
include:
- Calibration of gain control accuracy over a greater
range of output levels or instrument display indications than will
be used in any actual inspections to be performed.
- Checking instrument resolution under conditions
that do not duplicate inspection situations to be encountered.
- Measuring instrument output noise level under
conditions that have little connection with actual inspection conditions.
These potential shortcomings notwithstanding, it
might be argued that it is useful to employ some type of arbitrary reference
block to furnish standard targets for evaluating, on a periodic basis,
instrument sensitivity, resolution, and noise performance under fixed
conditions. Data so obtained would apply to a specific instrument only,
with a specific transducer continually coupled to the reference block.
Performance figures obtained for one instrument would not apply to another
instrument of different pulse shape, damping, and spectral response,
or to the results obtained from the same instrument used with another
transducer of different size, frequency response, damping, piezoelectric
coupling factor, orientation, and coupling to the block.
Because of the possibilities for changes in transducer
characteristics with time, and the variations that are possible in transducer
orientation and coupling efficiency (and even the possibility of changes
in block surface condition due to handling) it is doubtful that a given
block and transducer (contact or immersion) can furnish a very accurate
series of readings to be used for evaluating the stability of a given
instrument's characteristics over an extended time period, much less
its suitability for performing inspection of a particular material for
a particular type and size of discontinuity.
Despite these limitations, there is merit to requiring
a periodic instrumentation check. One benefit of periodic checking of
instrument performance is that it may reveal a gradual degradation in
certain areas that, while not serious enough to affect the ability to
detect discontinuities, as verified by calibrations taken before and
during inspection, does indicate the possible need for maintenance procedures
to be carried out. This could prevent future breakdowns or failures
to stay in calibration between checks made during inspection, either
of which could result in costly delays.
One method that has been used to overcome the difficulties
outlined above in obtaining a suitable means for long term evaluation
of changes in instrument performance consists of the construction of
a test block, as shown in Figure 1. This block has a permanently bonded
quartz transducer on one end and several side drilled holes at different
distances from the transducer end. The block is housed in a case to
protect it from external influences. It provides both an amplitude and
timing reference. This type of block has long been used to evaluate
several features of instrument performance for various types of instruments.
Although there is no way in which such a block can be traced to any
known NIST standard, it has proved to be a remarkably consistent device
for evaluating the stability of a given instrument's performance from
one calibration period to the next. No relationship is implied between
instrument performance so measured and its suitability to perform any
given inspection.
Figure 1 - Test block for
periodic instrument checks.
It may be desirable to perform periodic checks on
the amplitude linearity of an instrument even though accept or reject
decisions are made without reference to multiple amplitudes. Changes
in this characteristic may often reveal deterioration in electronic
components in the instrument that might ultimately lead to failures.
The accuracy specified for this characteristic should not exceed what
is needed for test validity or what is theoretically possible with the
checking method specified. It is impractical to do this reliably within
close tolerances by using the area amplitude characteristic of signals
obtained from flat bottom hole targets. This is made very clear by observing
that, even with the very specific transducer and instruments specified
for this purpose by ASTM E 127, that practice, based on careful
NIST studies, only specifies linearity as measured by standard ASTM
area amplitude blocks to be within +2, -3 dB accuracy limits (ASTM
E 127, 1997).
A superior method for measuring instrument linearity
is a calibrated attenuator (Bailey, 1977). ASTM Practice E 317
also provides for this method,which measures true instrument linearity
to a high degree of accuracy, is traceable to NIST standards through
the attenuator calibration, and removes the uncertainties and inaccuracies
due to variations in transducer coupling, orientation, wavelength, and
beam shape.
Appropriate Use of Flat Bottom Hole Reference
Blocks
There is no absolute correlation between the amplitude of the signal
response obtained by a given ultrasonic discontinuity detection instrument
from a flat bottom hole (FBH) target in a standard ASTM type reference
block and the signal that might be produced by the same instrument from
an actual discontinuity in material of another size, shape, and composition.
There appears, therefore, to be no justification for using such blocks
to set rejection criteria for the inspection of material of shapes,
sizes, and compositions other than that represented by the blocks themselves.
This point is explicitly made in ASTM Practice E 428 for steel
reference blocks in the section titled "Material Selection,"
and throughout the entire text of ASTM Guide E 1158 for material
selection for reference blocks.
Exceptions to the forgoing statement may be made
in cases where extensive successful use has generated enough data to
validate sensitivity calibration with a standard block to set rejection
criteria for certain types of discontinuities in certain specific materials
of other shape and composition when procedures are adhered to that limit
the characteristics of the transducers and instruments employed.
The most important use of these blocks lies in their application to
the measurement of the characteristics of one instrument or transducer
as compared to another under controlled, standardized (within limits),
and traceable target conditions. General sensitivity, resolution, and
noise performance can thus be assessed and used as a common basis for
instrument or transducer specification. These data must, however, be
carefully analyzed to determine their significance with respect to a
judgement of suitability for a given inspection application.
Conclusions
Requirements for periodic instrument or system qualification may often
be overstated in inspection specifications. As sometimes specified,
they can be impractical to perform, or worse, may be based on mistaken
assumptions about relationships among instrument, transducer, and material
properties. These requirements may also be misinterpreted as having
a direct bearing on the ability of a system to detect discontinuity
of a specified type in a specified material. The real use of qualification
checking should be to reveal long term changes in system performance
that might lead to future breakdowns or delays during calibration or
testing, or which might require frequent recalibrations to maintain
sensitivity during testing. For this purpose, care must be exercised
to specify checking by means that have long term stability and accuracy.
It must again be stressed that, whenever possible,
reference standards used to set sensitivity and rejection limits for
inspection of a given material should be fabricated from samples of
the same size, shape, and composition as the material to be examined.
There is also ample evidence to indicate that targets in reference blocks
should not be used to measure instrument linearity.
References
ASTM E 1158, "Standard Guide for Material Selection and Fabrication of
Reference Blocks for the Pulsed Longitudinal Wave Ultrasonic Examination
of Metal and Metal Alloy Production Material," Annual Book of
ASTM Standards, Vol. 03.03, 1997.
ASTM E 127, "Standard
Practice for Fabricating and Checking Aluminum Alloy Ultrasonic Standard
Reference Blocks," Annual Book of ASTM Standards, Vol. 03.03,
Section 11.4. 1997.
ASTM E 317, "Standard
Practice for Evaluating Performance Characteristics of Ultrasonic Pulse-Echo
Testing Systems Without the Use of Electronic Measuring Instruments,"
Annual Book of ASTM Standards, Vol. 03.03, Section 5.3.3. 1997.
ASTM E 428, "Standard
Practice for Fabrication of Steel Reference Blocks Used in Ultrasonic
Inspection," Annual Book of ASTM Standards, Vol. 03.03,
Section 4, 1997.
Bailey, J. A., "Ultrasonic Pulse-Echo Instrument
Linearity Measurement," Materials Evaluation, Vol. 35, No.
5, May 1977, pp. 64-68.
Beck, K. H., "Ultrasonic Area-Amplitude Limitations,"
Materials Evaluation, Vol. 50, No. 8, August 1992, pp. 978-985.
Gilmore, R.S., and G. J. Czerw, "The Use of
Radiation Field Theory to Determine the Size and Shape of Unknown Reflectors
by Ultrasonic Spectroscopy," Materials Evaluation, Vol.
35, No. 1, January 1977, pp. 37-45, 56.
Rogovsky, A. J., and J. L. Rose, "Additional
Aspects of the Standardless Sensitivity Selection Technique," Materials
Evaluation, Vol. 38, No. 9, September 1980, pp. 47-52.
Rogovsky, A. J., and J. L. Rose, "Ultrasonic
Sensitivity Selection Based on Probabilistic Techniques," Materials
Evaluation, Vol. 37, No. 4, March 1979, pp. 47-55.
- * Chair, Technical Instrument Corp., 152 Mercer County Airport,
Trenton, NJ 08628-1392; (609) 882-2894; fax (609) 882-3147; e-mail
tactices@aol.com.
Copyright © 1999 by the American
Society for Nondestructive Testing, Inc. All rights reserved.