Practical Issues in Electrical
Measurements on
Nanoscale Materials, Structures and Devices
by James
Niemann*
Figure 1-3
Figure
4-6
TESTING FOR UNIQUE PROPERTIES
Like the testing of other engineered
materials, measurements made on substances created with nanotechnology
are aimed at discovering useful properties. Generally, the relevant
properties relate to specific applications or products. For bulk
materials, the bulk properties of interest may include tensile
strength, phase transition temperature, weight, hardness and
electrical and thermal conductivity. Nano-technology offers ways to
create new materials that have improved properties for applications
such as structural members, electronic circuit elements, optical
devices and a host of other uses (Poole and Owens, 2003).
In some respects, the testing of bulk materials and those created
with nanotechnology are similar. However, the nature of nanotech
materials requires some novel testing techniques. Since these
materials are built at the atomic or molecular level, quantum
mechanics come into play. As a result of small particle sizes, the
atoms and molecules of these new materials may bond differently than
they might in bulk substances. There may be new electronic structures,
crystalline shapes and material behaviors. Nanoparticles with these
new properties can be used individually or as building blocks for bulk
material. While the discovery of bulk properties remains important,
measurements also need to uncover the characteristics unique to
nanoscale structures.
An excellent example is the carbon nano-tube. Carbon nanotubes can
be manufactured to consist of a single sheet of carbon atoms oriented
in a structure similar to graphite and rolled up into the shape of a
tube only a few nanometers in diameter (Collins, 1997). The smallest
diameter tubes are insulators (that is, large bandgap materials). As
the diameter of the carbon nano-tubes increases, its bandgap
decreases. With a large enough diameter, a carbon nanotube can become
a semiconductor. At even larger diameters, carbon nano-tubes exhibit
metallic properties, acting like conductors.
I
some respects, the testing of bulk
materials and those created with nanotechnology are similar.
STRUCTURE AND BEHAVIOR OF THE VERY SMALL
Particle size and structure have a major influence on the type of
measurement technique used to investigate a material. Macroscopic
materials can be viewed with optical microscopes. For nanoscopic
materials with particles smaller than 200 nm (7.9 ´ 10-6
in.), a scanning tunneling microscope or an atomic force microscope
can be used. Table 1 outlines the relative sizes of particles
considered as nanoscopic, mesoscopic and macroscopic.
| Table 1 Relative sizes of
various particle classifications |
|
| Classification |
Particle Size |
| Macroscopic |
larger than 300
mm
(1.2 x 10-5 in.) |
| Mesoscopic |
varies with phase
coherence length |
| Nanoscopic
|
0.5 to 200 nm
(2 x 10-8 to 7.0 x 10-6 in. |
| Atomic |
smaller than 0.5
nm (2 x 10-8 in.) |
|
Scanning electron microscopy, scanning tunneling microscopy and
atomic force microscopy techniques are particularly useful for
crystalline nanoparticle structures. Figure 1 illustrates the atomic
structures for some typical crystals with well known forms, such as
the simple cubic, body centered cubic and face centered cubic forms.
Knowing the arrangement of atoms in those structures helps predict a
particle's properties. However, on a nanoscopic scale, it is the
particle's size that fundamentally alters the physics of its behavior
and dictates the use of other measurement methodologies.
Two important sets of properties change as particle sizes are
reduced to nanometer dimensions: the material's chemical and
electrical characteristics. This even applies to biological materials
(Figure 2). Therefore, most of these materials require chemical and
electrical testing to characterize them for practical product
applications. For many of them, the actual quantity being measured is
a low level current or voltage that was translated from another
physical quantity (Grimnes and Martinsen, 2000). Direct electrical
measurements are possible on many nanoscopic substances, with probing
instruments and nanomanipulators now available.
Electronic Properties
As a substance is reduced to nanoscopic dimensions, both the bandgap
and the distance between adjacent energy levels within the material's
electron energy bands are altered. These changes, along with a
particle's nanoscopic size with respect to the material's mean free
path (average distance an electron travels between scattering events),
directly affect the electrical resistance of a nanoparticle. More
generally, a material's bandgap directly influences whether a particle
is a conductor, an insulator or a semiconductor.
These influential electronic properties allow, for example, a
carbon nanotube to be used to create a transistor switch (Martel et
al., 1998). One way to do this is by connecting a semiconducting
carbon document between two electrodes that function as a drain and
source. A third electrode (the gate) is placed directly under the
entire length of the carbon nanotube channel (Figure 3). For a
semiconducting carbon nanotube, the introduction of an electric field
through the channel (via the insulated gate placed in proximity to the
carbon nanotube channel) can be used to change the carbon nanotube
from its semiconducting state to its insulating state by increasing
the gate voltage. Decreasing the gate voltage will transition the
device into a conducting state (Figure 3b). This conduction mechanism
is analogous to the operation of a silicon mosfet transistor switch
(metal oxide silicon field effect transistor), which is created by
doping silicon with either an electron acceptor or donor to alter the
material's electronic conductivity in specific localities. (Additional
information on carbon nanotubes is available from the IBM T.J. Watson
Research Center [2005].)
Chemical Properties
The chemical properties of materials are also heavily influenced by
the electronic structure of the atoms within the molecules (bonding
type, bandgaps and so on) and by the size and shape of the particles.
Just as electronic properties change as particles are reduced to
nanoscopic dimensions, the bandgap and distance between adjacent
levels within electron energy bands alter chemical reaction
possibilities. In a substance composed of nanoscopic particles, a
larger percentage of the atoms in the mass will be exposed to the
material's surface. This also influences chemical properties,
especially for those molecules that can act as a catalyst.
PROPERTIES ASSOCIATED WITH ELECTRICAL MEASUREMENTS
For macroscopic particles, electrons take on discrete quantums of
energy that lie within energy bands, each band consisting of many
energy levels that electrons can share through their thermal energies.
For a conducting material, electrons can be thermally excited into the
conduction band - that is, electrons are present in the valence as
well as in the conduction band. For an insulator (where the bandgap is
greater than thermal energy of the electron), an enormous energy is
required for an electron to transition from the valence to the
conduction band separated by the material bandgap. If a suitable
amount of energy is absorbed (greater than the bandgap), then
electrons can jump bands.
As a particle's size is reduced to nanoscopic dimensions, the
allowable energies within the continuous bands separate into discrete
levels (since there are far fewer atoms in the mix). This occurs when
the separation between energy levels approaches the thermal energy of
the electrons (Figure 4). With fewer energy levels within the specific
energy band, the density of states of the material changes.
The density of states is a measure of the number of energy options
available to an electron as it falls into a lower energy level by
giving up energy or as it ascends to a higher energy level after
absorbing energy. A corollary is that if the density of states is
known, the size of the particle can be deduced.
The density of states can be used to engineer characteristics of a
nanoscopic particle, particularly a metallic or semiconducting
particle. For example, these materials take on color by selectively
absorbing and reflecting light wavelengths. If the material cannot
absorb photons with certain wavelengths because its density of states
has changed (removing critical energy levels associated with
absorption), then the color of the material will change.
If a particle becomes small enough, its physical size may approach
the wavelength of the material's electrons. Quantum size effects must
be considered whenever a particle approaches this critical dimension.
Because of quantum mechanics, the energy of its electrons cannot be
predicted by the bonding normally associated with the bulk material.
Nevertheless, electron energy effects can be deduced from
electrical measurements when nanoparticles take part in a chemical
reaction of the oxidation reduction type, such as the chemical
electrical conversion that takes place in fuel cells or batteries.
Briefly stated, every reaction of this type is associated with the
transfer of a specific number of electrons from one species to
another. The number of electrons transferred depends on the specific
reactions taking place and on the frequency of the reactions (reaction
rates). Electrical measurements of the transferred electrons can be
used to determine reaction rates by tracking the current and potential
of the cell, with time. Reaction and conduction measurements are then
used to track particle size, density of states, reaction rates and
other nanoscopic properties.
DETERMINING THE DENSITY OF STATES
Characterizing the density of states is a fundamental activity in
nanoscopic material research. Density of states (three dimensionality)
as a function of energy can be expressed as the number of electron
states per unit volume per unit energy at energy E
where
p(E) = the density of states per unit volume and per unit
energy
d = the partial derivative function
nS = the number of states (density of states) per
unit volume
m = the effective mass of the particle
h = Plank's constant
E = the energy (electron orbital location) in electron
volts.
While the result is independent of volume (can be applied to any
size particle), this equation is of limited value if the particle size
and structure are unknown. However, there are other ways to determine
the density of states experimentally from which the particle size can
be found. The 0D and 1D density of states as applied to nanoparticles
are discussed below.
X-ray Spectroscopy
The density of states can be measured by bombarding a material with
electrons. The energy of the bombarding electrons excites the material's
electrons into higher energy levels. X-ray emission takes place as
these electrons return to lower energy levels. The X-ray emission
energy indicates the difference between the energy level harboring the
excited electron and the level that recaptures the electron after the
emission. Fewer energy levels mean fewer electrons are excited,
yielding lower radiation intensities as they return to lower energy
states. A spectrographic analysis of the resulting X-ray emission
energies and intensities versus the bombarding electron energies
reveals the density of states for the material.
Direct Electrical Measurements
Since the density of states can be used to predict the electrical
behavior of materials, it is also possible to use electrical impedance
measurements to derive density of states information. Prior art has
used a scanning tunneling microscope to tunnel a current through a
nanoscopic device. The density of states is found by plotting
differential conductance versus applied voltage. Differential
conductance is simply di/dv.
In this technique, the quiescent current versus voltage
characteristics are established through the scanning tunneling
microscope's high resistance contact, with a low level alternating
current modulation on top of the quiescent operating point to measure
the differential conductance di/dv. When this conductance is
plotted against voltage, the graph indicates the material's density of
states.
Highly conductive materials possess an abundance of free energy
levels in the conduction band - that is, greater density of states
(more individual allowed energy levels per unit energy). Insulating
materials have an electronic structure with a dearth of occupied
energy levels in the conduction band. Since density of states
corresponds to the density of these energy levels, a plot of
conduction versus voltage provides a direct measure of the electronic
density of states at each energy level (voltage across the device).
Source Measurement Unit/Nanomanipulator Methodology
An alternate approach to this technique is to replace the scanning
tunneling microscope and its high resistance contact with a
nanomanipulator that makes low resistance contacts to the nanoparticle.
Such an arrangement allows charge transport and density of states
measurements without a scanning tunneling microscope. This works well
into the conduction region thanks to the low resistance direct
connections of the nanoprobes on the material (particle) being tested.
The nanomanipulator and its probes, along with a source measurement
unit, are used to apply a current or voltage stimulus directly to the
nanoparticle and measure its corresponding voltage or current response
(Figure 5). The advantage of electrical source measurement testing is
rooted in the fact that a specific source measurement unit measurement
mode (source current/measure voltage or vice versa) can be chosen
based on the relative impedance of the material or device under test.
Furthermore, the measurement mode can change dynamically as the
impedance changes, such as occurs in carbon nanotubes acting as
semiconductor switches. This allows a much wider dynamic range of
voltage and current stimuli and measurements, thereby optimizing
parametric test precision and accuracy. Source measurement unit
voltage and current sensitivity can be as good as 1 µV and 100 aA.
PROBING AND CONNECTION ISSUES
Electrical measurements on nanoscopic materials place stringent
requirements on the instrumentation. In order to measure conductivity,
impedance or other electrical properties and relate those measurements
to the density of states, a galvanic connection must be made to the
nanoscopic device under test (Keithley Instruments, 2004a). This
represents one of the major hurdles to be overcome in the field of
nano-technology testing. There are only a few tools available and few
device constructs that facilitate connections of this type.
Often, nanoscopic particles are classified dimensionally in terms
of exactly what it is about them that is small. Quantum wells
represent material with nanoscopic measurements in only a single
dimension, such as a thin film. Quantum wires are small in two
dimensions (or diameter), and quantum dots are nanoscopic in all
Cartesian coordinates.
Particle self assembly can be accomplished from silicon to silicon,
where conventional photolithographic techniques are used to make
electrical connection pads for probing. Particles that are long enough
to straddle such pads (for example, carbon nanowires) can be connected
to the pads through externally generated electrostatic fields. Figure
5 illustrates an example of this.
Although the properties of quantum wells, wires and dots differ, it's
possible that information about a particular material in the form of a
quantum dot can be inferred by examining the same material fashioned
as a quantum wire or well. Nanofilms are particularly easy to measure
since only one dimension is small. Such a film might be deposited on a
conductive substrate, allowing measurements through the volume as well
as over the surface, using appropriately placed macroscopic test pads
formed on the material surface. For conductive materials, separate
pads for source and measure can be deposited to create a kelvin (4
wire) connection (Keithley Instruments, 2004b). This type of circuit
eliminates test lead resistance from the measurement and improves
accuracy. In any case, a quantum well (nanofilm) can be tested like
any other bulk material.
A nanomanipulator and scanning tunneling microscope can also be
used to make direct electrical connections to nanoparticles. Scanning
tunneling microscopes do this by a tunneling current, established by
closing the distance between the scanning tunneling microscope tip and
the nanoparticle to within a single electron wavelength.
ELECTRICAL MEASUREMENT CONSIDERATIONS
Electrical measurements on passive devices (any device that is not a
source of energy) are made by following a simple procedure: stimulate
the sample in some way and measure its response to the stimulus. This
technique also works for devices that have both passive and active
properties with linear or nonlinear transfer functions. With
appropriate techniques, a source measurement algorithm can be useful
for characterizing sources of energy. Fuel cell and battery impedance
spectroscopy are examples of such measurements.
For nanoscopic particles, this general technique takes the form of
source measurement testing to quantify impedance, conductance and
resistance, which reveal critical material properties. This test
methodology is useful even if the end application is not an electronic
circuit.
General Considerations
Important considerations in the characterization of nanoscopic
particles include the following:
- Nanoscopic particles will not support the magnitude of currents
that macroscopic device can carry (unless they are
superconducting). This means that when a device is interrogated,
the magnitude of a current stimulus must be carefully controlled.
- Nanoscopic particles will not hold off as much voltage from
adjacent devices as a conventional electronic component or
material (such as a transistor). This is due to the fact that
smaller devices can and are placed closer together, smaller
devices have less mass and they may be affected by the forces
associated with large fields. In addition, internal electric
fields associated with nanoscopic particles can be very high,
requiring careful attention to applied voltages.
- Since nanoscopic devices are small, they typically have lower
parasitic (stray) inductance and capacitance. This is especially
useful when they are used in an electronic circuit, enabling
faster switching speeds and lower power consumption than
comparable macroscopic devices. However, this also means that
instrumentation for characterizing their current-voltage (I-V)
curves must measure low currents while tracking the short reaction
time.
Speed, Sensitivity and Accuracy
Since nanoscopic test applications often require low current
sourcing and measurement, appropriate instrument selection and use is
critical for accurate electrical characterization. Besides being
highly sensitive, the instrumentation must have a short response time
(sometimes referred to as high bandwidth), which is related to a
device under test's low capacitance and ability to change state
rapidly at low currents. This was illustrated earlier in Figure 3c.
Measurement Topology
It's important to recognize that the switching speed of a source
measurement test circuit may be limited by the instrumentation used to
follow the state of the device. This is especially true if a
nonoptimal measurement topology is used to observe the device. The two
possible topologies are source current/measure voltage, or source
voltage/measure current.
When considering the measurement of low impedance devices (less than
1
), the source
current/measure voltage technique will generally yield the best results.
Current sources are stable when applied to lower impedances, and a good
signal to noise ratio can be achieved without great difficulty. This
allows accurate low voltage response measurements.
The alternative, source voltage/measure current, is not as well
suited for low impedances. Exceptionally low values of applied voltage
are required to keep device currents low and avoid destructive
heating. At such low voltages, the source tends to contribute
excessive noise to the measured current (response). In other words,
the source's noise voltage is a significant percentage of the total
applied voltage. Additionally, voltage sources are less stable with
low impedance loads. There may also be current measurement problems
related to an instrument's voltage burden (the voltage that develops
across the input of an ammeter circuit), which introduces additional
error.
When measuring high impedance devices (impedance greater than 10
),
the source voltage/measure current technique is best. Stable
voltage sources to drive high impedances are easily constructed. When
a well designed voltage source is placed across a high impedance, it
will quickly charge the stray capacitance of the device under test and
test cables and rapidly settle to its final output value. The small
current response of the device under test can be accurately measured
with an appropriate ammeter.
The alternative, source current/measure voltage, creates problems
in high impedance measurements. To keep the voltage response low
enough for practical measurements, this technique requires a low
current value. This means that it will take a great deal of time to
charge the device and cable capacitances. In addition, the high
voltage measurement circuits will draw some of the source current from
the device under test. Since the current is sourced, not measured,
through the device, this current draw represents an error in the
measurement.
Electrical Noise
Measurement topology also has an effect on electrical noise, which
is the ultimate limitation on measurement sensitivity and accuracy.
For low impedance voltage measurements with a current source, the
measurement circuits will be sensitive to the device under test's
voltage noise and impedance.
For macroscopic devices, such as a resistor, the johnson noise
voltage at room temperature (293 K [68 ºF]) is expressed as:
where
k = Boltzmann's constant
T = absolute temperature of the source in kelvin
B = noise bandwidth in hertz
R = resistance of the source in ohms.
This can be further simplified to:
This equation shows that as device under test resistance R
decreases, the johnson voltage noise generated by the device under
test also goes down. Conversely, high impedance devices stimulated
with a voltage source are limited by current measurement noise. The
johnson current noise of a resistor at 293 K (68 ûF) is:
indicating that the noise goes down as device resistance increases.
For all particle sizes, in addition to johnson noise, there could
be a noise gain associated with the measurement topology chosen. Noise
gain is a parasitic amplification of the noise of the measurement
system that is not present when the correct measurement topology is
chosen. For example, consider a source voltage/measure current
topology. An operational amplifier is used in many current measurement
(ammeter) circuits, as shown in Figure 6. To minimize noise gain, the
ammeter circuit must operate at a low gain with respect to its
noninverting input terminal.
SOURCE MEASUREMENT INSTRUMENTS
A commercial direct current source measurement unit is a convenient
test tool for many nanoscopic material and device measurements. Source
measurement units change measurement topology automatically, that is,
rapidly switch between source voltage/measure current and vice versa.
This makes it easier to minimize measurement noise while maximizing
measurement speed and accuracy.
As described earlier for a carbon nanotube, some nanoparticles can
change state with the application of an external field. When
investigating such materials, a source measurement unit can be
configured to source voltage and measure current for a nanoparticle in
its high impedance state. When the material is in its low impedance
state, more accurate results are achieved by sourcing current and
measuring voltage. Furthermore, the source measurement unit has a
current compliance function that can automatically limit the direct
current level to prevent damage to the material or device under test.
Similarly, there is a voltage compliance function when current is
being sourced.
When using the compliance function, a source measurement unit will
satisfy the source value unless the user's compliance value is
exceeded. For example, when a source measurement unit is configured to
source voltage with a preset current compliance, if that compliance
value is exceeded, the source measurement unit automatically starts
acting as a constant current source. Its output level then will be the
compliance current value. Alternately, if the source measurement unit
is set to source current with a compliance voltage, it will
automatically switch to sourcing voltage (the compliance voltage) if
the device under test's impedance and the current it draws begin to
drive the voltage higher than the compliance value.
While a nanoscopic device, such as a carbon nanotube switch, can
change states rapidly, the change in instrument state is not
instantaneous. Depending on the source measurement unit model, the
switching time can range from 100 ns to 100 µs. Although such
switching speeds are not fast enough to track a nanoparticle as it
changes state, the time is short enough to allow accurate measurements
of both states while limiting device power dissipation to acceptable
levels.
Pulsing Techniques
Choosing the correct measurement topology to improve measurement
speed and minimize noise may still fall short of the test needs for
some nanoscopic materials. For example, it appears that some carbon
nanotubes can switch 1000 times faster than conventional complementary
metal oxide semiconductor transistor switches. This is too fast for
the nanoampere ranges of commercial picoammeters. Demanding devices
such as these may require other techniques to improve the speed of
impedance measurements.
Low power pulsing techniques may offer a partial solution to this
problem and are available in some source measurement unit designs. The
idea is to use a much higher test current or voltage and apply this
large stimulus for a short sourcing cycle. The larger stimulus will
lower the sourcing noise (by improving the signal to noise ratio) and
improve the rise or settle time for a voltage pulse or current pulse
respectively. Quieter sources require less filtering and permit a
shorter sourcing cycle time (narrower pulse width). A larger source
stimulus also increases the response current or voltage so that higher
instrument ranges can be used, further minimizing the effects of
noise. Since there is less noise, the measurement acquisition time
(integration period) can be shortened, thereby speeding up
measurements.
Avoiding Self Heating Problems
A possible source of error is self heating due to excessive
electrical current through the device under test. Such currents may
even lead to catastrophic failure of the sample. Therefore,
instrumentation must automatically limit source current during device
testing. Programmable current and voltage compliance circuits are a
standard feature of most source measurement unit based test systems
with pulsed current capabilities and may be required to avoid self
heating of some low resistance structures.
When an elevated test current is required, it must be sufficiently
brief so that it does not introduce enough energy to heat the device
under test to destructive temperatures. (Nanoscopic devices tolerate
very little heat, so the total energy dissipated in them must be
maintained at low levels.) In addition, care must be taken that the
magnitude of the test current is low enough that the device under test's
nanoscopic channel does not become saturated. (For instance, a current
channel 1.5 nm (5.9 x 10-8 in.) in diameter severely limits the
number of electrons that can pass per unit of time.) Some nanoscopic
devices can support only a few hundred nanoamperes of current in their
conductive state. Thus, a device's saturation current may define the
maximum test current even in pulsed applications.
The equation below illustrates how duty cycle and measurement time
in pulse mode effect device power dissipation. To calculate power
dissipation in pulse mode, multiply the apparent power dissipation (V
x I) by the test stimulus time and divide by the test repetition
rate:
where
Pp = pulse power dissipation
Pa = apparent power (V x I)
Tt = test time
Tr = test repetition rate.
Pulse mode is also useful for density of state measurements using a
low impedance connection, such as through a nanomanipulator. Pulsing
allows measurements at previously forbidden I/V locations due to
particle self heating.
OTHER INSTRUMENT ALTERNATIVES
Advanced alternating and direct current sources also offer a pulse mode.
This waveform generator permits the user to optimize pulse current amplitude,
pulse intervals, pulse width and trigger synchronization with a measuring
instrument, such as a nanovoltmeter (Keithley Instruments, 2004c). With
built in synchronization, the nanovoltmeter can take a reading within
microseconds after a pulse is applied. This greatly facilitates differential
conductance measurements and allows resistance measurements from 10
to 100
. Such an instrument combination is a high performance alternative
to alternating current resistance bridges and lock in amplifiers.
These advanced instruments can measure differential conductance up
to 10 times faster and with lower noise than earlier solutions. This
is accomplished in a single sweep, rather than by averaging the
results of multiple sweeps, which takes more time and is prone to
error. In addition, these instruments can be used in current reversal
mode, which further improves measurement accuracy by yielding device
impedance without the effects of thermally generated voltages. Taken
together, these techniques can improve measurement accuracy by as much
as three orders of magnitude compared to some test solutions.
ELECTRICAL IMPEDANCE SPECTROSCOPY
Electrical impedance spectroscopy is relevant to many devices that
utilize nanoscopic materials. One example is an electrochemical cell
that utilizes a nanoscopic material as a catalyst on a membrane
electrode assembly. The reactive component of the cell's complex
impedance provides a direct measure of the chemical reaction rate at
the terminals of the anode and cathode of the cell - which is a direct
reflection of the catalyst operation. In other applications, the
reactive component can describe the dielectric charge distribution of
a material and the ease or difficulty that an external field would
have in reorienting the material dielectric.
This technique goes beyond the simple measurement of impedance
magnitude with a direct current voltage or current. Complex impedance
is an alternating current (or pulsed direct current) property that can
be described as a vector with a magnitude and a phase. The phase
describes the relationship, in time, between the voltage or current
stimulus (zero phase reference) and the resultant current or voltage
response. At any frequency, the impedance can be described in this
manner as a magnitude with an associated phase angle. Complex
impedances must be computed at a specific excitation frequency; phase
angle changes with frequency for any device that can be modeled as a
resistance in series or parallel with capacitance or inductance.
By measuring complex voltage and current, the vector impedance can
be calculated by dividing the complex voltage by the complex current.
This requires only that the relative time is logged with each voltage
and current measurement. The final results are found by computing the
complex Fourier transform, which expands time domain data into the
frequency domain.
The complex impedance phase angle describes exactly the time lead
or lag between the stimulus and resulting device under test response,
since any phase angle can be transformed into a time by the following
relationship: phase/360 = time x frequency.
This is the shift in time between the stimulus and the response,
regardless of whether each is a voltage or a current. So, if we record
time along with voltage and current measurements, we can, through an
appropriate mathematical transform, compute the complex impedance.
We have already discussed the importance of selecting the correct
measurement topology to minimize noise and maximize system speed. In
addition, to accurately characterize complex impedance, the instrument
and measurement technique must allow an appropriate sample rate.
Furthermore, the instrument must have a stable time base in order to
compute the impedance mathematically. The required sample rate,
acquisition time and mathematical transform will depend on the
required accuracy and nature of the device's complex impedance. A
discussion of these criteria can be quite lengthy and is beyond the
scope of this tutorial. More information on digital signal processing
is available from Smith (1997).
APPLICATION EXAMPLE: CARBON NANOTUBE FIELD EMISSION DISPLAY
Typically, when carbon nanotubes are made, both conducting and
semiconducting forms occur. When the two forms are separated, the
conducting nanotubes can be used as emitters for field emission
displays and the semiconducting nanotubes can be used to make
electronic switches as discussed earlier (Figure 3).
Carbon nanotube field emitters play a role similar to that of the
electron gun used in a cathode ray tube. However, each pixel in a
field emission display has its own electron gun, a carbon nanotube
field emitter. Each emitter must be smaller than the display pixel and
able to emit electrons without dissipating too much heat. Since the
carbon nanotube acts as a cathode and this is a cold cathode emission
process, a very high electric field intensity is required. This is
possible because of the carbon nanotube's nanoscopic dimensions (Yao
et al., 2000).
The carbon nanotube field emitters must have other properties that
make this application practical. They must not erode away in the
presence of the pixel currents flowing through them. This means that
the emitter material's molecular bonding must be stronger than the
energies associated with the emission current. The physical
characteristics of the emitter must be such that a gate structure can
be placed between it (the cathode) and the phosphor coated screen of
the display (the anode). This gate is used to switch the pixel on and
off.
A field emitter display is a capacitive device, so the appropriate
measurement topology is the source current/measure voltage type. These
data are used to calculate the field emitters' impedance and
measurements are taken for emitter conductance over the entire
operating frequency range. Emitter voltage is measured from the common
cathode to the common anode, with one or many emitters enabled via
their addressable gates. Alternating current impedance can be measured
from direct current up to the maximum required refresh rate of the
display. Since emitters are electrical conductors, a low impedance
with acceptable capacitive and inductive components indicates a good
display. Naturally, a conductor such as this should have adequate
densities of states in the conduction band to support the required
conduction current.
Even though nanoscopic devices tend to have low capacitance due to
their small size, additional mechanisms must be considered when they
are combined with other materials and structures, such as those in a
field emission display. The macroscopic design of such a display
predicts a very high capacitance between the anodes, gates and
cathodes. Nevertheless, due to the dimensions required for cold
electron emission, each cathode current must be in the range of
nanoamps to microamps and must turn on or off quickly if the display
is to be used (and tested) as a raster device. This calls for
measuring low currents at a high bandwidth with the proper measurement
topology and techniques discussed earlier.
CONCLUSION
The electronic structure of nanoscopic particles is a reflection of
the atomic electron energies and the distribution of orbitals for both
molecularly shared and free electrons. This kind of information can be
used to describe how such materials will interact in the presence of
energy and other materials. The density of states in a material is
directly related to its electronic structure and is useful in
predicting or manipulating its properties. It can be found through
direct electrical measurements of differential conductance. Thus, the
density of states can predict a material's electrical impedance and
vice versa.
Still, there is a right way and a wrong way to electrically
interrogate a nanoscopic material, depending on its impedance. For a
low impedance material, the source current/measure voltage method will
result in the least electrical noise and allow the most accurate
response measurement with the widest bandwidth. For a high impedance
material, the source voltage/measure current method is more
appropriate for similar reasons. At times, the appropriate measurement
mode must be used in unison with yet another voltage or current source
used to activate or stimulate the device, such as in the case with the
field emission display.
REFERENCES
Collins, Philip G., "Nanotube Nanodevice," Science, Vol. 278, 30
October 1997, p. 100.
Grimnes, S. and O. Martinsen, Bioimpedance and Bioelectricity
Basics, London, Academic Press, 2000.
IBM T.J. Watson Research Center, <www.research.ibm.com/nanoscience/publications.html>, Yorktown
Heights, New York, accessed 2005.
Keithley Instruments, I-V Measurements of Nanoscale Wires and Tubes
with the Model 4200-SCS and Zyvex S100 Nanomanipulator, Application
Note #2481, Cleveland, Ohio, Keithley Instruments, 2004a.
Keithley Instruments, Four-probe Resistivity and Hall Voltage
Measurements with the Model 4200-SCS, Application Note #2475,
Cleveland, Ohio, Keithley Instruments, 2004b.
Keithley Instruments, Low Noise Pulsed I-V Measurements with the
Models 6221/2182A, Application Note Series, Cleveland, Ohio, Keithley
Instruments, 2004c.
Martel, R., T. Schmidt, H.R. Shea, T. Hertel and Ph. Avouris, "Single
and Multiwalled Carbon Nanotube Field-effect-transistors, Applied
Physics Letters, Vol. 17, No. 73, 1998, pp. 2447-2449.
Poole, Jr., Charles P. and Frank J. Owens, Introduction to
Nanotechnology, Hoboken, New Jersey, John Wiley & Sons, 2003.
Smith, Steven W., The Scientist and Engineer's Guide to Digital
Signal Processing, San Diego, California Technical Publishing, 1997.
Yao, Zhen, Charles L. Kane and Cees Dekker, "High-field Electrical
Transport in Single-wall Carbon Nanotubes," Physical Review Letters,
Vol. 84, No. 13, 2000, pp. 2941-2944.
* Keithley Instruments, Inc., 28775 Aurora
Road, Cleveland, OH 44139; (440) 248-0400; e-mail <niemann_james@keithley.com>.
Copyright © 2005 by the American Society for Nondestructive Testing, Inc. All
rights reserved.