Figure 1-3
General Description
of Computed Tomography
Computed
tomography (CT) is a radiographic inspection method that uses a computer
to reconstruct an image of a cross sectional plane of an object. In
conventional radiography, information on the slice plane P projects
into a single line, A-A; whereas in the associated CT image,
the full spatial information is preserved (see
Figure 1). The CT image is derived from a large number of systematic
observations at different viewing angles, and an image is then reconstructed
with the aid of a computer. If an internal feature is detected in conventional
projection radiography, its position along the line of sight between
the source and the film is unknown. Somewhat better positional information
can be determined by making additional radiographs from several viewing
angles and triangulating. This triangulation is a rudimentary, manual
form of tomographic reconstruction. In essence, a CT image is the result
of triangulating every point in the plane from many different directions.
Computed
tomography affords extensive capabilities for a variety of applications.
The systematic observations are
a set of X-ray attenuation measurements made along a set of paths projected
at different locations around the periphery of the test article.
Figure 2a illustrates a set of measurements made on a test object
containing two attenuating disks of different diameters. The X-ray attenuation
measurement made at a particular angle
is referred to as a single view. It is shown as f(x'), where
x' denotes the linear position of the measurement.
Figure 2b shows measurements taken at several other angles f(x').
Each of the attenuation measurements within these views is digitized
and stored in a computer where it is subsequently conditioned (for example,
normalized and corrected) and filtered (convolved). The next step to
create the CT image is to backproject the views, which is also shown
in Figure 2b. Backprojection
consists of projecting each view back along a line corresponding to
the direction in which the projection data were collected. The backprojections,
when enough views are employed, form a faithful reconstruction of the
object.
The resulting 2D cross sectional
image is a quantitative map of the linear X-ray attenuation coefficient
at each point in the
slice plane of the test article. The linear attenuation coefficient
characterizes the local instantaneous rate at which X-rays are removed
during the scan, by scatter or absorption, from the incident radiation
as it propagates through the object. The attenuation of the X-rays as
they interact with matter is a well studied problem and is the result
of several different interaction mechanisms. For industrial CT systems
with peak X-ray energy below a few megaelectronvolts, all but a few
minor effects can be accounted for in terms of the sum of just two interactions
photoelectric absorption and Compton scattering. The photoelectric
interaction is strongly dependent on the atomic number and density of
the absorbing medium; the Compton scattering is predominantly a function
of the electron density of the material. Photoelectric attenuation dominates
at lower energies and becomes more important with higher atomic number,
while Compton scattering dominates at higher energies and becomes more
important at lower atomic number. In special situations, these dependencies
can be used to advantage.
One particularly important property
of the total linear attenuation coefficient is that it is proportional
to material density, which is a fundamental physical property of all
matter. The fact that CT images are proportional to density is one of
the principal virtues of the technology and the reason that image data
are often thought of as representing the distribution of material density
within the object being inspected. However, this can be an oversimplification.
The linear attenuation coefficient also carries an energy dependence
that is a function of material composition. The energy dependence of
the attenuation coefficient may or may not be more important than the
density dependence, depending on the materials and the energies of the
X-rays involved. Once the CT system is calibrated for density measurements
(an algorithm for density calibration is defined in ASTM document
E1935 Standard Test Method for Calibrating and Measuring CT Density),
the linear attenuation coefficient of an unknown feature in an image
can be measured from a determination of its mean CT value. Its density
can then be extracted from a knowledge of its mass attenuation coefficient.
Types of Computed Tomography
Scanners
Computed tomography scanners are identified by the mechanical equipment
configuration which provides the relative motion between the test article,
the source, and the detectors. It makes no difference, at least in principle,
whether the test object is moved systematically relative to the source
and detectors or if the source and detectors are moved relative to the
test object. Physical considerations such as the weight or size of the
test article should be the determining factors for the most appropriate
motion to use.
The majority of scan geometries
that have been employed can be classified as one of five generations.
This classification is a legacy of the early, rapid development of CT
in the medical arena, and these terms are still widely used. The various
scan geometries are illustrated in
Figure 3.
First generation CT systems (Figure
3a) are characterized by a single X-ray source and single detector
that undergo both linear translation and rotational motions. The source
and detector assembly is translated in a direction perpendicular to
the X-ray beam. Each translation yields a single view, as shown in
Figure 2. Successive views are obtained by rotating the test article
and translating again. The advantages of this design are simplicity,
good view to view detector matching, flexibility in the choice of scan
parameters (such as resolution and contrast), and ability to accommodate
a wide range of different object sizes. The disadvantage is a longer
scanning time.
Second generation CT systems (Figure
3b) use the same translate/rotate scan geometry as the first generation.
The primary difference is that second generation systems use a fan beam
of radiation and multiple detectors so that a series of views can be
acquired during each translation, which leads to correspondingly shorter
scan times. Like first generation systems, second generation scanners
have the inherent flexibility to accommodate a wide range of different
object sizes, which is an important consideration for some industrial
CT applications.
Third generation CT systems (Figure
3c) normally use a rotate only scan geometry, with a complete view
being collected by the detector array during each sampling interval.
To accommodate objects larger than the field of view subtended by the
X-ray fan, it is possible to include part translations in the scan sequence,
but data are not acquired during these translations as during first
or second generation scans. Typically, third generation systems are
faster than their second generation counterparts; however, because the
spatial resolution in a third generation system depends on the size
and number of sensors in the detector array, this improvement in speed
is achieved at the expense of having to implement more sensors than
with earlier generations. Since all elements of a third generation detector
array contribute to each view, rotate-only scanners impose much more
stringent requirements on detector performance than do second generation
units, where each view is generated by a single detector.
Fourth generation CT systems (Figure
3d) also employ a rotate-only scan motion. The difference between
third generation and fourth generation systems is that a fourth generation
CT system uses a stationary circular array of detectors and only the
source moves. The test specimen is placed within the circle of detectors
and is irradiated with a wide fan beam which rotates around the test
article. A view is made by obtaining successive absorption measurements
of a single detector at successive positions of the X-ray source. The
number of views is equal to the number of detectors. These scanners
combine the artifact resistance of second generation systems with the
speed of third generation units, but they can be more complex and costly
than first, second, or third generation machines. Furthermore, they
require that the object fit within the fan of X-rays, and they are more
susceptible to scattered radiation.
Fifth generation CT systems (Figure
3e) are different than the previous modes in that there is no mechanical
motion involved. The scanner uses a circular array of X-ray sources,
which are electronically switched on and off. The sources project onto
a curved fluorescent screen, so that when an X-ray source is switched
on, a large volume of the part is imaged simultaneously, providing projection
data for a cone beam of rays diverging from the source. This method
of data collection is essentially different from the other four, since
a series of 2D projections of a 3D object is collected rather than a
series of 1D projections of a 2D object. This scanning mode is appropriate
for precise imaging of a rapidly moving part (typical application is
imaging of heart or other moving organs).
Tradeoffs in CT Instrumentation
Currently, the selection of a CT system geometry is application specific.
The generational differences are more or less suited to certain areas.
A significant factor in driving medical CT systems to use rotate-only
scan geometries was the requirement that scanning times be short compared
to the length of time that a patient can remain motionless or that involuntary
internal motion can be ignored (that is, seconds). These considerations
are not as important for industrial applications in which scan times
for specific production related items can typically be much longer (that
is, minutes) and the dose to the object is often not an important factor.
A second generation scan geometry is attractive for industrial applications
in which a wide range of part sizes must be accommodated, since the
object does not have to fit within the fan of radiation as it generally
does with third or fourth generation systems. A third generation scan
geometry is attractive for industrial applications in which the part
to be examined is well defined and scan speed are important. To date,
first, fourth, and fifth generation scan geometries have seen little
commercial application, but there may be special situations for which
they would be well suited.
Advantages and Disadvantages
of CT Technique
The principal advantage of CT is that it nondestructively provides quantitative
densitometric (that is, density and geometry) images of thin cross sections
through an object. Because of the absence of structural noise from detail
outside the thin plane of inspection, images are much easier to interpret
than conventional radiographic data. And, with proper calibration, dimensional
inspections and absolute density determinations can also be made very
accurately.
As with any modality, CT has its
limitations. The most fundamental is that potential objects for examination
must be small enough to be accommodated by the handling system of the
CT equipment available to the user and must be radiometrically translucent
at the X-ray energies employed by that particular system. Further, CT
reconstruction algorithms require that a full 180 degrees of data be
collected by the scanner. Another potential drawback with CT imaging
is the possibility of artifacts in the data. As used here, an artifact
is anything in the image that does not accurately reflect true structure
in the part being inspected. Because they are not real, artifacts limit
a user's ability to quantitatively extract density, dimensional, or
other data from an image. Therefore, as with any technique, the user
must learn to recognize and be able to discount common artifacts subjectively.
Common Industrial Applications
of CT
Computed tomography provides a nonsuperimposed visual image of the internal
structures of a part and is fundamentally a nondestructive inspection
tool. However, the industrial applications where CT has proven most
valuable are in the areas of rapid prototyping, reverse engineering
and metrology.
Rapid prototyping can be accomplished
utilizing a class of manufacturing techniques where parts are built
from computer models in a variety of materials. Stereolithography is
one such technique that can utilize the information of CT to produce
extremely accurate polymer parts. Taking multiple CT slices, the 2D
images can be assembled to produce complete 3D representations of scanned
components. The data is presented to the stereolithography system as
full volume information or simply contour plots, allowing the generation
of either filled or hollow polymer parts. The choice of data would be
based on the rapid tooling techniques that are applied in the specific
application area.
Computed tomography assisted reverse
engineering methods are successful in enabling older designs without
computer aided design (CAD) files to access the many available rapid
tooling techniques currently available. In reverse engineering applications,
as in rapid prototyping, the 2D images can be assembled to produce complete
3D representations of scanned components. There are many computational
methods that allow the CT derived digital data to be transformed to
CAD contours, which can be used to reverse engineer the part. The CAD
contours produced from CT data have been determined to be accurate to
within a few thousandths of an inch. Thus, CT data is similar to dimensional
data from coordinate measuring machines except it provides a number
of advantages CT data is acquired without contacting the part, CT
data not only provides surface information but also accurate measurements
of all internal structure, and CT images can be formed of any object
without special programming, regardless of its structural complexity.
Metrology of the CT data evaluating
dimensional measurements can be accomplished using a number of techniques.
The most accurate means is by reverse engineering the slice data. This
method requires the generation of a point cloud a collection of points
in 3D space that represent the surface of the part from the CT data
and registering that with the CAD model of the part. The deviations
between the inspection data and the design data are evaluated based
on the necessary tolerances for the application.
Computed tomography affords extensive
capabilities for a variety of applications. In addition to a nondestructive
inspection technique for quality control, CT data is a record of the
different material densities present in the part and feeds into other
programs to provide rapid prototyping and reverse engineering capabilities.
Conclusion
Computed tomography is a radiographic method that provides an ideal
examination technique whenever the primary goal is to locate and size
planar and volumetric detail in three dimensions. Because of the relatively
good penetrability of X-rays, CT permits the nondestructive physical
and, to a limited extent, chemical characterization of the internal
structure of materials. Also, since the method is X-ray based, it applies
equally well to metallic and nonmetallic specimens, solid and fibrous
materials, and smooth and irregularly surfaced objects. The ability
of a CT system to image thin cross sectional areas of interest through
an object makes it a powerful complement to conventional radiographic
inspections, and, when used in conjunction with other NDT methods, such
as ultrasound, CT data can provide evaluations of material integrity
that cannot currently be provided nondestructively by any other means.
References
Herman, Gabor T., Image Reconstruction from Projections: The Fundamentals
of Computerized Tomography, New York, Academic Press, 1980.
Newton, T.H., and D.G. Potts,
Eds., Radiology of the Skull and Brain, Vol 5: Technical Aspects
of Computed Tomography, C.V. Mosby Company, 1981.