Microradiography
as Strong NDT Tool
by
Bahman Zoofan*
|
The magnifying capabilities of conventional
X-rays tubes are limited due to focal spot sizes in the range
of millimeters. Magnification of more than two to three times
can result in unsharp, fuzzy images. Microfocus tubes were developed
with focal spot sizes measuring tens of micrometers; this was
the birth of microradiography. Even though today's techniques
are reasonably applied to predominantly thin or low absorbent
material, drastic advances in digital radiography are expected.
Dietmar Henning
Contributing Editor |
Figure 1-4
Figure
5-7
Figure 8-10
INTRODUCTION
It has been reported that when
Wilhelm Roentgen discovered X-rays, it took only one month to
determine most of their properties. From an early time, radiographers
have been interested in the resolution of the radiographic image. Due
to the relatively large focal size of conventional radiographic
systems, geometrical unsharpness significantly limits the image
resolution. An analogy for this effect is the size of a light source
and the sharpness of the resultant shadow. In the very early years of
radiographic development with X-rays, some literature referred to the
technique as shadowgraphy, to describe image formation by
electromagnetic radiation.
In an attempt to produce magnified images with X-rays, two English
engineers, C.T. Heycock and F.H. Neville, tried to generate magnified
images by passing the beam of X-rays through a lead collimator with a
fine hole (Poen, 1959).
The setup that they used, which is shown schematically in Figure
1, is limited in its effectiveness as it only allows a small
portion of the X-ray beams to pass through and it is not very
practical to make fine round holes in thick lead.
In
recent years, microfocal systems have become more commercially
available as an NDT tool.
The need to observe fine details in different applications, such as
microjoining, electronics, materials testing, biology and medicine,
led radiographers to the development of techniques that optically
enlarge high resolution radiographs. Figure
2 shows an example of a small electronic part taken with a
conventional tube without any magnification by R.C. McMaster around
1960 at the Ohio State University. He used a very large source to film
distance to make the radiograph as sharp as possible. The radiograph
was enlarged by printing it on photographic paper with dimensions of
203 by 254 mm (8 by 10 in.). Using this technique, he was able to
detect fine porosity in soldered joints, which would be very difficult
to resolve with a conventional image. This work illustrates the
benefits that can be achieved by enlarging high resolution images and
thus set the direction for the development of microradiography.
The need for high resolution X-ray imaging techniques posed
challenges for NDT engineers and pushed the design and manufacture of
affordable microfocus tubes. It is likely that technical progress in
this field was helped by the manufacture of small powerful vacuum
systems and reliable electronic control units.
To see the effect of the source size on the resolution of X-ray
images, we should look at the basic concept of geometric unsharpness
in radiography.
Considering the geometric unsharpness Ug in a radiographic image
with an effective source size F, source to detector distance a and
source to object distance b then:
| (1) |
|
Defining the image magnification M, like any other imaging
technique as:
| (2) |
|
Then the geometric unsharpness (Equation 1) can be written as:
| (3) |
|
This simple relation shows that in order to enlarge the
radiographic images with enough sharpness, the source size should be
in the range of a few micrometers. A simple classification of X-ray
tubes considers the focal spot sizes between 2.5 to 5 mm (0.1 to 0.2
in.) as large, less than 2.5 mm (0.1 in.) as small and less than 100
µm (3.9 x 10-3 in.) as fine (recently tubes with 0.5
µm [2 x 10-5 in.] focal spot size have been advertised).
Figure 3 shows the effect of
the source size on the geometric unsharpness of the radiographic
image. Figure 3a shows a
magnified image with conventional radiography using a large focal spot
size. The geometric unsharpness in this case is unreasonably large and
completely destroys the clarity and fine detail within the image. The
recommended technique for reducing the unsharpness with a large focal
spot radiographic source is to place the object and the detector (film
in this case) in close contact as shown in Figure
3b. Figure 3c illustrates
how a fine source can be used for projection magnification and still
result in negligible unsharpness in the produced radiograph.
Figure 4 shows the geometric
unsharpness versus focal spot size for different magnifications (based
on Equation 3). From this figure, it is clear that large
magnifications are only practical when fine focal spot sizes are used.
To obtain an image with a geometric unsharpness of less than 0.2 mm
(7.9 x 10-3 in.) an X-ray source with a focal size smaller
than 100 µm (3.9 x 10-3 in.) should be used. The
selection of the optimum projection magnification for a specific
source size depends on different conditions and needs more discussion.
The use of a microfocal system has been traced back to 1960 (Fontijn
and Boogerd, 1969), but the applications were limited to some research
centers (Boogerd et al., 1976; Fontijn, 1975; Parish and Cason, 1977).
In recent years, microfocal systems have become more commercially
available as an NDT tool. In general, the capability and real
potential of microradiography is not well known by many industrial
radiographers who are used to using conventional radiography. This is
partly because few microradiography publications are available and
most codes and radiographic standards recommend that the geometric
unsharpness should be kept as small as possible by positioning the
sample as close as possible to the detector or film.
Therefore, the concept of projection magnification, in which the
object is far from the detector, is unfamiliar to many professional
radiographers. Figure 5 shows
examples of sharp, enlarged radiographs taken with a microfocus tube.
The other advantage of microradiography is that by keeping the
specimen a long way away from the detector, the scattered radiation
generated inside the object does not reach the detector. This
reduction in the scattered radiation can improve the clarity of the
image dramatically when compared to conventional radiography. Figure
6 shows an enlarged real time microradiographic image of a US dime
with 4x magnification. The high contrast and clarity of the image
reveal the fine surface contours on both sides of the coin.
The high resolution images obtained by microradiography with a
minimal amount of scattered noise can be further magnified optically
to improve the detectability of details, as shown in Figure
7. This microradiograph shows fine shells used for geological
research.
DIFFERENT FEATURES OF MICROFOCAL TUBES
Hairpin Filament
One of the main differences between a conventional tube and a
microfocus tube is the design of the filament (cathode) inside the
tube (Figure 8). In order to
have a fine focal spot in the tube, the beam of electrons that hits
the target should be as fine as possible while keeping an acceptable
energy level. To achieve a fine beam of electrons, filaments are bent
into a hairpin shape (Figure
8c). The material of the cathode is usually tungsten, with wire
thickness of about 0.2 mm (7.9 x 10-3 in.). Other types of
materials are sometimes used for special applications.
Demountable Tube Head
A consequence of using a hairpin filament in microfocus tubes to
produce a fine electron beam is the short life, which requires the
cathode to be replaced after only limited hours of operation. The
microfocus tube heads are therefore designed for plug-in replacement.
Since the seal and vacuum are broken during replacement, it is also
necessary to reestablish vacuum before reenergizing the tube.
Typically, microfocal systems have a small and efficient vacuum system
consisting of a rotary pump in conjunction with a turbomolecular pump.
The volume of the tube head is an important factor in the design of
microfocal systems and can be tuned for a normal working environment
so that evacuation only takes a few minutes. Another important factor
in the design of a microfocus tube head compared to a conventional one
is having a short distance between the target and the tube window.
This distance should be in the range of a few millimeters, compared to
several centimeters for a conventional one. This distance is important
because the object should be close to the tube window to achieve a
large projection magnification. As is apparent, all these factors in
designing a tube head for a microfocal system present many engineering
challenges.
Electron Beam Focusing
After the operational vacuum is reached inside the tube, the electron
beam can be accelerated by the applied high voltage at the anode that
helps the electrons pass through an opening and emerge as a fine
accelerated beam toward the rotary target.
The focusing unit is an electromagnetic, asymmetrical electron
lens. The centering unit contains the electromagnetic equipment for
correcting any deviation of the electron beams with respect to the
tube axis.
Smart Control Unit
It can be imagined that a microfocus tube with so many features needs
a smarter controller when compared to a conventional unit. A control
unit typically indicates the status of the vacuum, automatic warmup,
filament condition, setup for kilovoltages and applied tube current
and exposure time.
The control unit and associated computer programs also control the
output power and automatically expand the focal spot size if the
applied power is increased beyond its setting to prevent burning the
target. The control unit continuously indicates whether the X-ray tube
is in the microfocal mode or not. Usually fiber optics are used for
fast and reliable communication between the control unit and the
generator.
Microfocal systems for Failure Analysis
The combination of a microfocal tube with a real time imaging system
and precise mechanical manipulation equipment with multiple axes to
accurately position objects in front of the X-ray tube is a powerful
NDT tool, particularly for failure analysis purposes. Figure
9 shows the results of a test on a safety switch typically used to
prevent overcharging of an electronic device.
Phase Contrast Imaging
One of the interesting features of microfocal systems is their ability
to create phase contrast images. This new and promising imaging
technique needs fine focus tubes or X-ray radiation with high spatial
coherence. This imaging technique can be used to give enhanced images
of low absorbent materials, which inherently produce low contrast
radiographs (Zoofan et al., 2005). The technique is based on small,
refracted angles and phase shifts in the X-rays after passing through
an object that has a density gradient. Figure
10 shows an example of a phase contrast image taken by a 5 µm
(2 x 10-4 in.) microfocal system.
Some limitations of microfocal systems should be mentioned here as
well. One of the first limitations is their inherent low output
wattage. The applied current in a microfocal system is in the range of
microamperes, compared to milliamperes in conventional tubes. The fact
that microdiscontinuities are more critical for fine structures with
thinner thicknesses makes microfocal systems more practical for
testing of thin, low absorbent materials. Electronic components,
biological samples, plastics and parts with moderate thicknesses of
aluminum and titanium are great candidates for microradiography. The
second limitation of microfocus tubes is their high price compared to
conventional units. The higher price range often prevents their use in
companies with limited capital resources. Maintenance and costly
repair can be considered another limitation of these systems. The last
major limitation of a microfocal system is the lack of practical and
widely accepted techniques for the measurement of exact focal spot
size. Customers generally have to accept the claim of the tube
manufactures when they refer to the focal spot size of their
microfocal systems.
Due to the fact that microradiography is a relatively new imaging
technique, there is no related code or standard available. The finest
1 T hole in thin penetrameters can be magnified and made visible by
this technique. Even test patterns with 10 lp/mm (254 lp/in.) are not
fine enough to check the performance of a system. The best way to
evaluate the resolution of a microfocal system is by using fine
resolution patterns with details in the range of a few micrometers,
used for scanning microscopy.
Despite its limitations, microradiography is becoming a powerful
NDT tool and is continuing to be innovatively applied in areas where
traditional radiography cannot meet the needs of industry, academia
and research.
REFERENCES
Boogerd, W.J., P. van Zuylen and L.A. Fontijn, "An Unconventional 150
kV Microfocus X-ray Equipment for NDT Purposes," British Journal of
NDT, Vol. 18, November 1976, pp. 175-178.
Fontijn, L.A., "Mini-magnetic Lenses for Microfocus X-ray
Applications," Journal of Vacuum Science and Technology, No.
12, 1975, pp. 1359-1362.
Fontijn, L.A. and W.J. Boogerd, "New Demountable 0.1 mm X-ray Tube,"
Metal Construction, Vol. 1, No. 12, 1969, p. 585.
Parish, R.W. and D.W.J. Cason, "High Definition Radiography of Cast
Turbine Blades As a Method of Detecting and Evaluating the Incidence
of Microporosity," NDT International, August 1977, pp. 181-185.
Poen, O.S., Microprojection with X-rays, Dordrecht, The
Netherlands, Martin Nijhoff, 1959.
Zoofan, B., S.I. Rokhlin and G.S. Frankel, "Application of Phase
Contrast Microradiography in NDT," Materials Evaluation, Vol.
63, 2005, pp. 1122-1127.
* The Ohio State University, 190 Edison, 1248
Arthur Adams Dr., Columbus, OH 43221; (614) 292 3012; e-mail
<zoofan.1@osu.edu>.
Characterization Sciences Group, Pacific Northwest
National laboratory, PO Box 999, MSIN: K5-26, Richland, WA 99352.
Copyright © 2005 by the American Society for Nondestructive Testing, Inc. All
rights reserved.