ADVANTAGES AND DISADVANTAGES
Digital Photography
There are a number of variables that can affect the utility
of using digital photography to capture images of penetrant indications
(Figure 1). To digitally photograph fluorescing crack indications, consideration
must be given to the resolution of the camera. For instance, is a 1.3
megapixel resolution sufficient or is 5 megapixels the minimum? Another
issue that arises when using digital images is that of image processing.
Image processing software can be used to manipulate the image before
printing. Hue, contrast and saturation can all be adjusted, but what
effect does this have on the integrity of the image? Remember that the
goal is to create as realistic an image as possible, not to create an
idealized version of the indications.
Other factors also have to be considered. The type of
paper (matte or glossy) used to print photographs may affect their utility.
The resolution of the printer is also important. Images taken at high
resolution are going to be of limited value if printed on a low-resolution
printer. There are additional factors (discussed below) that may affect
how one views the resulting images.

Figure 1 - Digital photo (6 Mb) of calibration panel for comparison
purposes.
Film Photography
To capture images of fluorescing crack indications on
film, consideration must be given to the resolution and quality of the
camera lenses, as well as paper type and quality. Since the required
image must include both the fluorescent emissions from the cracks as
well as the surrounding background, some visible light must be present
to illuminate the chrome surface. At the same time, reduced light levels
must be maintained to achieve maximum contrast of the fluorescent indications.
With both digital and film photography, the final image
size needs to be carefully controlled for the inspector to be able to
usefully compare the photo with the actual cracked panel.
Finally, while ultraviolet-A light is required to view
the calibration panel, white light is required to properly view a photograph.
Inspectors also require some time for their eyes to adjust to the new
lighting. Although it is possible to view photographs under ultraviolet-A
lighting, the indications on the photograph will not fluoresce, so the
comparison is not "like with like."
Opaque Adhesive Tape
Opaque tapes have been used for years as a means of lifting
developer, and the indication contained with it, from the surface of
the test object. This can then be stored as is or pressed onto clear
acetate film to prevent movement of the indications. It must be kept
in mind that this requires the use of more developer than might ordinarily
be applied. It also requires careful consideration of storage and viewing
options.
Images stored in this way are viewed under ultraviolet-A
illumination either directly on the adhesive surface containing the
developer/indications or through the acetate sheet. If the image has
been pressed onto acetate film, there is a potential loss of fluorescence
due to the absorption of ultraviolet radiation by the acetate.
Transparent Adhesive Tape
The use of clear adhesive tape is similar to that of opaque
tape in that more developer is required. Clear tape has the advantage,
however, that it can be fixed to an opaque base (such as paper or a
card) and the indication viewed through the tape. Again, the absorption
of the ultraviolet illumination by the tape must be taken into account
when selecting a suitable medium.
Finally, since these adhesive tape approaches take the
actual fluorescent penetrant from the surface, they are subject to the
same problems encountered with the dyes in the in-use penetrant, in
that they can fade when repeatedly exposed to ultraviolet-A illumination
(and, of course, sunlight) and therefore may require regular replacement.
Storing in dark conditions and viewing for the minimum time aids in
the protection of the images.
Strippable Lacquer with Developer
Strippable coatings, by their very name, imply that removal
from the test surface is easily achieved. This is not always the case,
and a measure of patient practice is required if all the coating is
to be removed from rough cast or shot-blasted surfaces. It is here that
an application of developer can be of assistance. The action is similar
to the use of "parting powder" in sand casting, where the
application is used to assist in the separation of the cope and drag
and the removal of the pattern.
However, the prime function of the developer is to create
negative (or reverse) capillarity and cause the penetrant to creep out
of the discontinuity and spread over the surface. This results in the
enlargement of the indications, with the potential to create a replica
whose indications appear larger than those on the discontinuity standard
itself. The use of developer should therefore be limited to exactly
the same procedure as for the daily performance test - usually an application
of dry powder developer in a dust storm cabinet.
Finally, the replica is an actual sample of the penetrant
contained in the indication, and therefore is subject to the same fading
problems as with adhesive tape.
Strippable Lacquer without Developer
Applying the coating directly on the surface reduces the
spread of the indications but does not completely eliminate it. It should
also be borne in mind that the image produced by lacquer replica is
a mirror image. As an aid to stripping the replica, a somewhat thicker
coating may be required, but like the developer-assisted replica, it
still suffers from the problems of fade and embrittlement with age.
Plastic encapsulation can be used, but the type of plastic needs to
be carefully chosen, as some types absorb ultraviolet-A.
Direct Measurement (Overall Area)
The actual measurement of the starburst requires ultraviolet-A
illumination, a suitable, calibrated loupe (or measuring graticule)
and a clear understanding of what is to be measured (Figure 2). This
is not the sort of operation that would be undertaken as part of the
regular performance test, but is a very useful guide for determining
changes that could make the panel unsuitable for use.
Measuring the overall area of the individual starburst
requires a graticule on which is etched a series of circles. These are
available and are used primarily for sizing gas holes, gas pores and
porosity in castings. The circle that is able to contain the whole of
the starburst is taken to represent the size of that starburst. These
results are then recorded and reassessed on a regular basis, the frequency
of which will depend upon the process specification requirements.

Figure 2 - Individual starbursts (6 Mb digital image) for size change
evaluation.
Direct Measurement (Maximum Dimension)
Once again, ultraviolet-A illumination, a suitable, calibrated
loupe (or measuring graticule) and a clear understanding of what is
to be measured are required. Measuring the maximum dimension of the
individual starburst requires a graticule on which is etched a linear
scale. These are available and are used primarily for sizing cracks,
laps and linear indications. The maximum dimension of the starburst
is taken to be the longest leg of that starburst. These results are
then recorded and reassessed on a regular basis, the frequency of which
will depend upon the process specification requirements.
The advantage of both these approaches is that the individual
crack is being measured and no actual replication is required. This
eliminates any enlargement problems as well as the fading and embrittlement
issues.
Results
There is no single approach for assessment and replication
of known discontinuity standards. Currently, all of those described
above are specified in one form or other by a variety of contractors,
leaving some choice to the supplier, but since each approach has its
own advantages and disadvantages, a clear choice for a single approved
approach is not obvious.
Conclusion
Conclusions on this topic are not easy to draw. While
there are advantages and disadvantages to each of the various approaches
I've discussed above, none of the possible solutions to the problem
at hand stand out as definitively superior to the rest.
This is where I hope that the reader will come in. Any
comments would be gratefully received, not just from the prime contractors
who create the requirements but also from the inspectors who must apply
these approaches and the Level IIIs who must interpret the requirements
for their own company procedures. I can be reached at the e-mail address
given in the contact information on the first page of this article,
and I look forward to hearing from you.