|
Back to Basics [ click here for the Back to Basics Archive ]
Volumetric X-Ray Testing
by Harold Berger*
and Robert L. Schulte*
|
As radiographic techniques become more
sophisticated to solve more sophisticated problems, the basics
of the technique become more complicated. Here is a great
explanation of how one of the new imaging techniques, tomography,
works. Berger and Schulte give a really short and to the point
explanation - and without the pages of math usually found in
such articles. The photographs of how it works are worth many
thousands of words that would never have done the job - at least
for me.
Frank Iddings
Tutorial Projects Editor
|
Figure 1-3
Figure 4-6
Introduction
Most radiographers have come across an
X-ray testing
problem that was not solved by a conventional two dimensional X-ray image. The
information sought was hidden behind other structures in the object. If the
system was a realtime radioscopic system, the information might be obtained (or
might not, depending on the geometry) by rotating the object in the beam while
the test was being conducted. However, many installed X-ray testing systems,
even realtime radioscopic systems, don't have the necessary object rotation
capability. In these cases, the desired information may be obtained by a
repositioning of the object in the beam, perhaps several times, to see the
desired X-ray image detail. An alternative to this approach is a volumetric
X-ray imaging system. Volumetric X-ray image techniques have existed for some
time (Ellingson and Berger, 1980), but many X-ray people will think first about
computed tomography (ASTM International, 1997; Bossi, 1996). However, there are
many testing situations where the size, shape or location of the object is not
suitable for testing using computed tomography methods.
Images of any region
of a testing object can be obtained quickly and easily
with volumetric X-ray systems.
There are various terms that have been used to describe
the process of creating a useful X-ray image of a selected plane within an
object. Planigraphy and laminography are some of the older terms used for this
process of bringing a selected plane in the object into good focus. The
technique was used for many years in the medical community (Plantes, 1932). In
the simplest form, the object remained stationary while the X-ray source was
moved linearly in the +X direction and the imaging detector (usually film) was
moved at the same time and rate in the -X direction. This resulted in a single
image plane in the object (the fulcrum plane) remaining in focus while images
from other object planes were blurred along the direction of the source detector
movement. The process could be repeated to image other object image planes. The
motion for the source detector movement may also be more complicated than the
simple linear motion described; the blurring of out of plane detail is more
effective when more complex source detector movements are used. For example,
circular or hypocycloidal motions provide images of the desired object plane
with less blurring from out of plane object details.
The terms used to describe this process for computer
methods are "tomography" and "tomosynthesis." Tomography is defined in ASTM
E-1316 (1999) as follows: "any radiologic technique that provides an image of a
selected plane in an object to the relative exclusion of structures that lie
outside the plane of interest." Tomosynthesis is normally used to describe a
technique in which the X-rays are directed through the object in a limited
number of views, with the objective of reconstructing multiple image planes
within the object.
Computed tomography is widely used in the medical
community and is receiving increased attention from industrial users. Computed
tomography systems are usually configured to take many views of the object,
often more than 100; this provides reconstructed images of good quality, with
excellent density discrimination. Limited view tomosynthetic techniques are
beginning to attract attention from a number of investigators, many of whom are
taking advantage of new, semiconductor digital imaging detectors - particularly
flat panels (Antonuk, 1995; Jones and Berger, 2000; Jones, 2001). Many reports
on advances in computed tomography and tomosynthesis X-ray imaging are described
in recent symposium proceedings (Antonuk and Yaffe, 2001; Boone and Dobbins,
1999; Dobbins and Boone, 2000). These advanced techniques have also been
described at the recent series of ASNT topical meetings on digital imaging (for
example, see American Society for Nondestructive Testing, 2001).
An example of an advanced volumetric X-ray testing
technique is an all electronic, digital, multiimage technique that permits an
operator to reconstruct any horizontal or vertical X-ray image plane in the
object (Schulte et al., 2001; Schulte 2001; Schulte, 2002). This capability
allows the viewer to look behind obstructing details that may hide the testing
region of interest. The operator also has the capability to scan through
selected regions of the object in both the horizontal and vertical directions.
Since an integral part of this system is an immediate response (realtime
radioscopic capability), the operator can view the usual two dimensional X-ray
image and decide if more information is needed to complete the test. If that is
the case, additional images can be taken while the testing object is in place to
provide the volumetric images needed to complete the test.
Volumetric Imaging System
A simple system for stereo imaging requires only two
images taken from angles that permit stereo image viewing. X-ray image stereo
viewing can be accomplished with film or filmlike detectors (American Society
for Nondestructive Testing, 1985; Evans and Robinson, 2000) or in realtime
radioscopy by rapid movement of the
X-ray focal spot in a microfocus X-ray tube (Polansky et
al., 1990). However, more than two X-ray views are required for reconstruction
of multiple X-ray image planes within an object. True tomosynthesis X-ray
systems for viewing of multiple image planes require more than two X-ray images
and provide much more image information. The system, like all tomosynthesis
techniques, makes use of multiple two dimensional X-ray images taken from
different perspectives.
The tomosynthesis process requires a series of projection
X-ray images taken from different angles through the test object. A schematic of
the technique is shown in Figure 1. (Jones and Berger, 2000; Schulte et al.,
2001; Schulte, 2001). A typical testing geometry is shown with the source
located at an oblique angle with respect to the vertical direction. The object
under test is rotated to eight different positions and an
X-ray image is acquired at each position. The process is
very flexible and is not constrained to any specific positions, number of images
or geometry. The circular image pattern, however, is convenient to demonstrate
the technique. The image detector can be any flat detector, film, scintillator
or flat panel. The flat panel, as used in this system, provides rapid response,
excellent dynamic range and digital images for input to the reconstruction
software. In this case, eight two dimensional digital images are collected,
showing details in the testing object from the different image perspectives. Any
view through the object can be reconstructed with software using the basic set
of oblique incidence images acquired during the X-ray test.
After acquiring the limited X-ray image data set, the
application software is used to reconstruct views of any horizontal X-ray image
plane parallel to the flat panel image detector face. In addition, the software
can display any vertical X-ray image plane (perpendicular to the flat panel
face). The reconstructed X-ray image planes provide an excellent dimensional
measurement capability with the accuracy dependent on the resolution of the
image detector. An interesting feature of the application software is a realtime
scan through the X-ray image planes. This dynamic display of internal
X-ray images is very useful. It allows an operator to look
rapidly through the X-ray image volume of the object, stop at any point and
reconstruct the entire X-ray image plane at that level. This dynamic response
can't be shown in a hard copy article, but several examples of reconstructed
X-ray images are shown in the following section.
Applications
A tomosynthetic examination of a test object can be used
to illustrate the power of this technique. Two aluminum plates were prepared -
each 12.7 mm (0.5 in.) thick plate contains a pair of flat bottom 3.175 mm
(0.125 in.) diameter holes at various depths. The plates were assembled into a
single 25 mm (1 in.) thick object with the hole openings facing the interface
between the two plates. Reconstructed horizontal X-ray images of each plate show only the holes in that
plate. Figure 2 shows horizontal views at two different levels within the
assembled plate object. Figure 2a shows a horizontal X-ray image view at a level
of 10 mm (0.4 in.) above the detector plane, well within the lower aluminum
block (two holes are shown). Figure 2b, taken at 15 mm (0.6 in.), shows only the
other two holes in the upper block. A cross sectional view of the plates can be
obtained through a vertical image reconstruction. Figure 3 shows the horizontal
view at the interface of the two plates at which each of the four holes is just
visible. Using the application software, a multisegment line is drawn over all
the holes, as is shown in Figure 3. A vertical reconstruction is made along the
line segments, starting in the upper right corner (Figure
4). The image is a
cross section of the stacked plates showing all of the holes as a function of
depth; the holes in each plate extend only to the interface between the plates.
The diameters and depths of all the holes can be measured with precision on the
order of 127 µm (5 x 10-3 in.).
The test object described above illustrates what can be
done with this tomosynthesis X-ray imaging system. A practical application of
X-ray tomosynthetic imaging is weld testing. The horizontal and/or vertical
X-ray reconstructed images provide excellent location information about weld
discontinuities. This, coupled with the capability for dimensional measurement,
gives an inspector the information needed to decide about repair and from which
side of the weld to start. Figure 5 shows a horizontal X-ray image
reconstruction through a butt weld of two 12.7 mm (0.5 in.) thick aluminum
plates with a single V groove. The total thickness of the weld is 18 mm (0.7
in.). This horizontal image, at a level of 16 mm (0.63 in.) is just below the
level of the plates, showing discontinuities (a pore and inclusions) in the weld
bead. A vertical X-ray image (Figure 6) reconstructed along the direction of the
weld and through the center of the pore shows the size and location of the pore.
This information makes it easier to assess the quality of the weld.
Conclusion
Full volumetric X-ray images provide additional
information for the detection and characterization of discontinuities or
components in all kinds of structures, including welds, castings, electronic
devices and electromechanical assemblies. Use of a digital flat panel provides
high sensitivity, fast response and good resolution. Images of any region of a
testing object can be obtained quickly and easily with volumetric X-ray systems.
This includes systems such as the electronic X-ray system described in this
article. Compared to other volumetric X-ray image systems, the X-ray system
offers advantages, including reconstructed viewing of any horizontal or vertical
X-ray image plane, scanning through selected horizontal or vertical image
regions and a precision measurement capability. Volumetric X-ray imaging, using
methods such as tomosynthesis or computed tomography, can provide the additional
information often needed for critical nondestructive testing applications.
Acknowledgments
The authors wish to thank Donald Twyman, of the Digitome
Corporation, for his assistance with the preparation and review of the paper.
References
American Society for Nondestructive Testing, Digital
Imaging IV, Columbus, Ohio, ASNT, 2001.
American Society for Nondestructive Testing,
"Radiographic Flaw Depth Determination," Nondestructive Testing Handbook,
second edition: Volume 3, Radiography and Radiation Testing, L.E. Bryant and P.
McIntire, eds., Columbus, Ohio, ASNT, 1985, pp. 808-816.
Antonuk, L.E., "Flat-panel Horizons in Digital X-Ray
Imaging," Photonics Spectra, Vol. 29, No. 6, 1995, pp. 108-116.
Antonuk, L.E. and M.J. Yaffe, eds., Medical Imaging 2001:
Physics of Medical Imaging, Proceedings of SPIE, Vol. 4320, 2001.
ASTM International, ASTM E-1441, Standard Guide for
Computed Tomography (CT) Imaging, West Conshohocken, Pennsylvania, ASTM
International, 1997.
ASTM International, ASTM E-1316, Standard Terminology for
Nondestructive Examinations, West Conshohocken, Pennsylvania, ASTM
International, 1999.
Boone, J.M. and J.T. Dobbins, III, eds., Medical Imaging
1999: Physics of Medical Imaging, Proceedings of SPIE, Vol. 3659, 1999.
Bossi, R.H., "Radioscopy and Tomography," Nondestructive
Testing Handbook, second edition: Volume 10, Nondestructive Testing Overview, S.
Ness, C.N. Sherlock, P.O. Moore and P. McIntire, eds., Columbus, Ohio, American
Society for Nondestructive Testing, 1996, pp. 173-198.
Dobbins, J.T., III, and J.M. Boone, eds., Medical Physics
2000 - Physics Medical Imaging, Proceedings of SPIE, Vol. 3977, 2000.
Ellingson, W.A. and H. Berger, "Three-Dimensional
Radiographic Imaging," Research Techniques in Nondestructive Testing, R.S.
Sharpe, ed., London, Academic Press, Vol. 4, 1980, pp. 1-38.
Evans, J.P.O. and M. Robinson, "Design of a Stereoscopic
X-Ray Imaging System Using a Single X-Ray Source," NDT&E International,
Vol. 33, 2000, pp. 325-332.
Jones, T.S. and H. Berger, "Performance Characteristics
of an Electronic Three-Dimensional Radiographic Technique," ASNT Spring
Conference and 9th Annual Research Symposium Abstracts, Columbus, Ohio, ASNT,
2000, pp. 29-31.
Jones, T.S., "Evaluation of Digital X-Radiological
Imaging Systems for US Air Force Applications," Materials Evaluation, Vol. 59,
2001, pp. 971-975.
Plantes, Ziedes des, "Eine Neu Method zur Differenzierung
in der Roentgenographie," Acta Radiologica, Vol. 13, 1932, pp. 182-192.
Polansky, D., T. Jones, R. Placious, H. Berger and J.
Reed, "Real-Time Stereo X-Ray Imaging," 1990 ASNT Spring Conference - NDE:
the Next Decade and Beyond, Columbus, Ohio, ASNT, 1990, pp. 193-195.
Schulte, R.L., T.S. Jones and H. Berger, "A Digital X-Ray
Inspection System for Full Volumetric Imaging," Digital Imaging IV, Columbus,
Ohio, ASNT, 2001.
Schulte, R.L., "Full Volumetric Imaging Using a Digital
Flat Panel," ASNT Fall Conference, Columbus, OH, October 2001.
Schulte, R.L., "Full Volumetric Digital Imaging of
Welds," American Welding Society, Nondestructive Testing of Welds Conference,
Houston, Texas, American Welding Society, May 2002, pp. 195-208.
* Industrial Quality, Inc., 640 E. Diamond Ave., Suite C,
Gaithersburg, MD 20877; (301) 948-2460; fax (301) 948-9037; e-mail <hberger@indqual.com>.
Copyright ©
2002 by the American Society for Nondestructive Testing, Inc. All
rights reserved.
[ Materials
Evaluation ]
Copyright © 2008 by the American Society for Nondestructive Testing, Inc. ASNT is not responsible for the authenticity or accuracy of information herein. Published opinions and statements do not necessarily reflect the opinion of ASNT. Products or services that are advertised or mentioned do not carry the endorsement or recommendation of ASNT.
IRRSP, NDT Handbook, The NDT Technician and www.asnt.org are trademarks of the American Society for Nondestructive Testing, Inc. ACCP, ASNT, Level III Study Guide, Materials Evaluation, Nondestructive Testing Handbook, Research in Nondestructive Evaluation and RNDE are registered trademarks of the American Society for Nondestructive Testing, Inc.
ASNT exists to create a safer world by promoting the profession and technologies
of nondestructive testing.
|