by Quincy Howard and Steve Enzukewich*
|
While heat treatment does not typically
affect the velocity of acoustic waves, it definitely can effect
the acoustic attenuation, as seen in this month's NDT Solution.
The lesson learned is that reference standards with appropriate
heat treatment are necessary for some alloy steels to match
the complete acoustic response of the material.
G.P. Singh
Associate Contributing Editor
|
Figures 1-3
Figures 4-6
Introduction
Reference standards are used in ultrasonic
testing to calibrate the instrument (screen range, gain setting, etc.)
prior to performing an inspection. In the aerospace industry, it is
common practice to use reference standards that are specific to the
inspection to be performed, as opposed to a general standard like the
IIW block. These reference standards are designed to have the same material
and geometric properties as the part to be inspected, and are commonly
manufactured by an airline operator or a local machine shop. Since the
heat treatment condition of the material does not greatly effect the
sound velocity it is usually not necessary to specify a heat treatment
condition.
Problem
Several instances of excessive ultrasonic sound attenuation in materials
used to manufacture reference standards have come to our notice at Boeing
and through reports from airline customers. The instances have all been
alloy steel materials (4330, 4340, etc.) that were being tested at 10
MHz. Figures 1 and 2
are two instrument screen displays of a 10 MHz longitudinal beam transmitted
in two apparently identical 76 mm (3 in.) long alloy steel samples.
Figure 1 is the expected result
with no signals between the initial pulse and the backwall signal at
80 percent of screen width. In Figure
2, however, it is not possible to distinguish which signal is the
back surface reflection because of all the noise signals.
Sound attenuation as severe as in Figure
2 makes instrument calibration impossible. Even less severe sound
attenuation is a concern as it can cause very high instrument gain settings
which in turn can lead to false calls during the inspection. Alloy steel
airplane parts, which are fully heat treated, do not have these noisy
characteristics. A reference standard with these high noise characteristics,
therefore, does not truly represent the characteristics of the parts
to be inspected.
Another way to limit the amount of sound attenuation in reference
standards is to match them acoustically to the part to be inspected.
Possible Solutions
One solution to the situation above is to require that alloy steel reference
standards be heat treated to the "on-airplane" condition.
A full heat treatment would match the reference standard's metallurgical
characteristics with the part to be inspected and thus prevent any excessive
sound attenuation. This treatment consists of two high temperature heating
steps (austenitizing and tempering) separated by a rapid cooling (quenching)
step. Full heat treatment, however, is costly and many small machine
shops do not have the facilities necessary for proper heat treatment
of alloy steels. Another difficulty with a full heat treatment is the
possibility of geometric changes caused by the quenching step.
Another way to limit the amount of sound attenuation
in reference standards is to match them acoustically to the part to
be inspected. Two simpler heat treatments exist for alloy steels: normalizing
and subcritical annealing. Although these two heat treatments will not
match the exact microstructure of a fully heat treated part, they can
produce a material that has similar ultrasonic results.
These two heat treatments are used together to prepare
alloy steels for machining. Normalizing is used to homogenize the part
material (i.e., to create a uniform microstructure) before it is annealed.
Normalizing takes a material at some unknown microstructural state and
brings it to a known starting condition. The annealing step is done
after normalizing to soften the material prior to machining. It is called
a subcritical anneal when it is performed at temperatures below the
point at which austenite forms for the particular steel alloy. These
two heat treatments offer an advantage over a full heat treatment because
they involve only one heat cycle, and the cooling is accomplished in
air as opposed to rapid quenching.
The important question is whether either (or both)
of these heat treatments will produce a microstructure that is similar
to a fully heat treated part in its ultrasonic characteristics. In the
past, we have used normalizing of some reference standards in instances
where excessive sound attenuation difficulties have occurred, but a
direct comparison to a fully heat treated part (or subcritical annealed
part) has not been performed.
Experiment
An experiment was designed to compare these three different heat treatments.
We exposed 24 samples of "noisy" alloy steel material to one
of three different heat treatments (a full heat treatment, normalizing,
or a subcritical anneal). The samples were taken from three different
developmental reference standards. The heat treatments were performed
according to Boeing process specifications in a vacuum furnace to prevent
scaling of the surface. Some of the fully heat treated samples were
tempered to the 1,030-1,170 MPa (150-170 ksi) range and the others to
the 1,520-1,650 MPa (220-240 ksi) range to get a cross section of fully
heat treated values.
The relative sound attenuation was measured before
and after the heat treatments by comparing the amount of instrument
gain needed to get an 80 percent of full screen height backwall signal.
In many of the samples it was not possible to distinguish the backwall
signal in the initial state because of the excessive noise signals.
In these cases the gain settings were estimated based on the known screen
range and the expected location of the backwall signal.
The same ultrasonic flaw detector, transducers,
and instrument settings were used for all of the readings. Gain settings
on all of the samples were taken by the same two individuals and averaged.
Gain settings for each sample were taken at both 10 MHz and 5 MHz. A
3.2 mm (0.125 in.) diameter element, longitudinal wave transducer was
used for the 10 MHz readings. This transducer was selected because it
is used in one of the inspections where difficulties with sound attenuation
were first noticed. A 6.25 mm (0.25 in.) diameter element, longitudinal
wave transducer was used for the 5 MHz readings.
Three of the samples were also sectioned and examined
metallurgically to determine what caused the excessive sound attenuation
and what the heat treatments did to reduce it. These samples were all
taken from the same original piece so that microstructure comparisons
would be of the same original material.
Results
Prior to any heat treatment, all of the samples at 10 MHz showed excessive
sound attenuation. In all cases it was very difficult to determine with
certainty which signal was the back surface reflection. Figure
3 is an example of a typical screen display. It took an average
of 88 dB of instrument gain to get an 80 percent of full screen height
signal. At 5 MHz, however, it took an average of only 63 dB to get the
same signal. The 5 MHz signals were very "clean" with no excess
noise signals present.
As expected, the fully heat treated samples showed
the most improvement from the original samples. Figures 3
and 4 show the same sample
at 10 MHz before and after the full heat treatment respectively. The
full heat treatment removed all of the excess noise as can be seen in
Figure 4 and made distinguishing
the backwall signal very straightforward. The improvement in the samples
at 10 MHz ranged from 32.5-40 dB after the full heat treatment was performed.
There was very little change in the sound attenuation of the samples
when tested at 5 MHz. Because of the relatively low dB starting point
at 5 MHz, the samples improved only 4.3-8 dB after the full heat treatment.
The samples that were normalized showed improvements
very similar to the fully heat treated samples both at 10 MHz and 5
MHz. The samples improved 24.9-33.2 dB at 10 MHz after normalizing.
Figures 5 and 6
show one of the samples at 10 MHz before and after the normalizing,
respectively. The screen display shown in Figure
6 is very similar to that for the fully heat treated sample shown
in Figure 4.
The subcritical annealed samples showed little or
no improvement at either 10 MHz or 5 MHz. In the case of the 5 MHz readings,
more gain was needed after the subcritical anneal to get the same 80
percent screen height signal. The relatively low temperatures of the
subcritical annealing process did not make a significant enough microstructural
change to affect the sound attenuation readings.
Table 1 summarizes the sound attenuation results for all of the samples
at both frequencies. The greatest improvement can be seen in the 10
MHz readings; little change can be seen at 5 MHz. On average it took
8 dB more at 10 MHz to get the same signal in the normalized samples
than it did in the fully heat treated samples. In both cases the signals
were very clean, with no noise signals between the initial pulse and
the backwall signal.
Three samples, representing the untreated condition,
the fully heat treated condition, and the normalized condition were
cross-sectioned, etched, and examined under 250 magnification. The untreated
sample showed a microstructure typical of an as-cast condition (no subsequent
mechanical or thermal treatments) which is very inhomogenous in both
grain shape and size. There were several very large areas of pearlite
(a mixture of ferrite and plate-like carbide) present in the microstructure.
At high ultrasonic test frequencies (and therefore short wave lengths),
these areas, if they are sufficiently large, can become reflectors of
the sound. This explains why the noise signals were present at 10 MHz
and not present at 5 MHz. The presence of these relatively large pearlite
areas and the general inhomogenous condition of the material at the
grain boundaries both could have contributed to the high sound attenuation
seen in these samples.
The fully heat treated sample showed a very fine
and homogenous grain structure. The uniformity of the grain structure
and the very small grain size would account for the relative ease in
which the sound traveled through the samples. The normalized sample
showed a very homogenous grain structure as well, but with grains many
times larger than the fully heat treated sample. Even though the grains
were larger than those of the heat treated sample, they were made up
of ferrite (pure iron) and not the plate-like pearlite. A quick test
was made at 15 MHz to see if the larger grain size of the normalized
samples had any sound attenuation effect at a higher frequency. There
was no noticeable difference between the normalized and fully heat treated
samples which indicated the larger grain size of the normalized samples
did not affect the sound transmission.
Table 1 Comparison of experimental
results
| |
Samples
at 10 MHz |
Samples
at 5 MHz |
| Full heat
treatment |
Average of 37 dB
improvement |
Average of 6 dB
improvement |
| Normalized |
Average of 29 dB
improvement |
Average of 4 dB
improvement after normalizing |
| Subcritical
anneal |
Little to no effect |
Slightly negative
effect |
Conclusions
Heat treating an alloy steel ultrasonic reference standard to the "on
airplane" condition is the best way to match the standard to the
part. Reference standards in this condition best represent the part
to be inspected, both from a metallurgical standpoint and an ultrasonic
standpoint.
The tests described in this paper showed that normalizing
alloy steel reference standards is a simple and relatively inexpensive
way to match the acoustic characteristics of alloy steel reference standards
to those of a fully heat treated part. Even though a normalized microstructure
is not the same as a fully heat treated microstructure, normalizing
will remove unwanted noise signals and requires only one short heating
cycle followed by cooling in air. Although gain settings appear to be
slightly higher with normalized reference standards compared to fully
heat treated ones, the gain settings will not be high enough to cause
excessive noise signals or false calls.
The tests showed that a subcritical anneal does
not improve noise signals in alloy steel reference standards. The temperature
for subcritical annealing is too low to change the microstructure enough
to remove unwanted noise signals.
The tests also showed that the excessive sound attenuation
in these samples was very dependent on the instrument/transducer frequency.
In the longitudinal mode, normalizing was not necessary when inspecting
the samples at 5 MHz. (Because of the slower sound velocity of shear
and surfaces waves, and the subsequent shorter wave lengths, sound attenuation
may be a difficulty when inspecting at 5 MHz in these modes.) Other
alloy steel materials may have other microstructural conditions where
5 MHz longitudinal wave inspections encounter excessive sound attenuation.
Acknowledgments
The authors thank Brian Pearce for the "noisy" test piece
materials, Norm Munsey for the heat treating, and Luther Gammon for
the metallography.
* The Boeing Company, (425) 342-3164; fax (425) 342-0665; e-mail quincy.howard@boeing.com.
Copyright © 1997 by the American
Society for Nondestructive Testing, Inc. All rights reserved.
[ Materials Evaluation
]