Figures 1-3
Introduction
Calibration
blocks are used to assess the performance of ultrasonic flaw detection
equipment, including transducers. They are also used to calibrate the
equipment (setting up of gain, range, sweep rate, etc.) for the examination
of materials. The blocks are supposed to be made of materials having
a specified composition and heat treatment, and of a given surface texture
and geometrical form. Variation in any characteristic from block to
block may lead to different interpretation of results. It is desirable
that all the calibration blocks, used anywhere in the world, have the
same general properties.
While it is not difficult to verify the dimensions
and the geometry of blocks, the test for other properties such as composition,
heat treatment, and surface texture is not straightforward. These other
properties cast appreciable influence on ultrasonic velocity and attenuation.
Fortunately, these two ultrasonic parameters are desired to be nearly
the same in blocks, and if these parameters have the same values, one
need not go to the extent of evaluating composition, grain size, surface
texture, etc.
The ultrasonic velocity can be evaluated by any of
several methods having an accuracy of 0.1 percent or better. The evaluation
of ultrasonic attenuation, however, poses several problems. The transducer
characteristics (diameter, frequency, bandwidth, acoustic impedance of
facing material), couplant characteristics (thickness, acoustic impedance),
velocity in sample (diffraction correction), etc., can affect the measurement
of attenuation. Fortunately, the task is not so much to determine the
absolute value of attenuation but to check the variation in the attenuation
value from block to block. A method of evaluating the relative attenuation,
or a parameter depending on the relative attenuation, should therefore
serve the purpose very well. The method, of course, should be simple and
provide results having dependence on very few variables and yet have an
integrated influence of grain size, composition, and surface texture.
A change in any of these characteristics should change the result.
Variation in any characteristics from block to block may lead
to different interpretation of results.
The American Society for Testing and Materials has
described a method for the quality evaluation of aluminum standard reference
(cylindrical) blocks (ASTM, 1989). However, no method for the quality
assurance of another widely used calibration block VU (also known as
the IIW or A2 block) has been given in the literature available (IS,
1990; BS, 1978). In the method described by ASTM, a steel ball of 4.8
mm (0.18 in.) diameter is kept at a distance of 150 mm (6 in.) from
the transducer while the distance between transducer and cylindrical
block is kept at 50 mm (2 in.).
The height of the echo from the ball is used to
set up the gain of the flaw detection equipment. If the ball echo height
is 50 percent full signal height and the noise level between interface
echo and first backwall echo is 10 percent full signal height, the block
is said to be of acceptable quality. It is also desired that five backwall
echoes should remain observable while scanning the block.
In the present paper, a method is described which
in essence is similar to the method described in the ASTM method but
is simpler and independent of the quality of ball and water. It also
gives the quality evaluation of the material of the V1 block.
Theory
In the case of a large plate (such as a V1 block) the acoustic pressure
amplitude of first interface echo can be given as
(1) 
and that of seventh backwall echo as
(2) 
where P0 = initial pressure amplitude,
= attenuation coefficient
for water,
= attenuation coefficient
for material of the block, Lw = water path, D
= diffraction correction term for first interface echo, D
= diffraction correction term for seventh backwall echo, R =
reflection coefficient, T = transmission coefficient, and Lm
= thickness of the block.
Dividing P1 by P2
(3) 
The diffraction correction term D does not depend
upon block material and hence can be treated as the same for all the
blocks. The term in the denominator (DT2R8)
is dependent on transducer characteristics and acoustic impedance of
block material, besides other factors. The effect of transducer characteristics
can be obviated if the same type of transducer is used for the evaluation
of all the blocks. The variation in acoustic impedance of the material
from one block to another is usually very small and its effect on the
ratio P1/P2 will be insignificant as compared
to the effect of am on P1/P2. In other
words, maximum change in the ratio P1/P2
will be due to the change in attenuation of the material.
Taking Equation 3 into consideration and by some
mathematical manipulation, we can define the relative attenuation (RA)
as given by the term
(4) 
This can be read from the ultrasonic flaw detector
directly as the difference in echo heights in dB.
In the case of cylindrical blocks (ASTM, 1989),
though the expressions for the pressure amplitudes of the echoes will
change, the discussion remains unaltered.
Recommended Procedure
The block diagram for the experimental procedure is shown in Figure
1. The block under test is immersed in a water tank. An immersion
transducer of 5 MHz frequency, 10 mm (0.75 in.) diameter is placed in
a position where the interface echo from the front face is of maximum
height.
In the case of the V1 block, the transducer is moved
to a distance where the second interface echo falls halfway between
the seventh and eighth backwall echoes (Figure
2). This nearly fixes the distance between transducer and reflector.
The first interface echo is brought to 50 percent full signal height
and gain is noted in dB. Then the seventh backwall echo is brought to
the 50 percent full signal height by increasing gain of the flaw detector
and new gain is noted in dB. The difference in gain in dB is the required
relative attenuation.
In the case of the cylindrical reference blocks
(ASTM, 1989), the transducer is kept at a distance of 150 mm (6 in.)
to avoid the beam profile effects of the transducer (Figure
3). The transducer is positioned to get maximum height of the interface
echo. The relative attenuation is evaluated using first interface echo
and second backwall echo, taking both at 50 percent full signal height.
To avoid confusion because of the presence of the echoes due to mode
conversion, higher backwall echoes are not considered.
Acceptance Levels
On the basis of results obtained from several blocks, we find that in
the case of the V1 blocks made of low carbon alloy steel having grain
size of Mcquaid-ehn No. 5, the relative attenuation should be 35 dB.
In the case of the cylindrical block made of aluminum alloy (DTD, 1973),
the relative attenuation should be 26 dB. A block having relative attenuation
more than the above specified value may not be acceptable, while one
within 2 dB down may be acceptable. It may be noted that the above acceptance
levels have been arrived at by the experimental observations for specific
materials. In case the recommendations are for different material, the
acceptance level has to be worked out.
Acknowledgment
The authors thank Professor E.S.R. Gopal, Director of the National Physical
Laboratory, New Delhi, for taking interest in this work. They are also
grateful to Om Prakash Vaish and Jagdish Lal for rendering help at various
stages during the experiment.
References
ASTM E 127: Recommended Practice for Fabricating and Checking Aluminum
Alloy Ultrasonic Reference Blocks. ASTM, 1989.
BS 2704: Calibration Blocks for Their Use in
Ultrasonic Flaw Detection. British Standards
Institution, 1978.
DTD 5124: Aerospace Material Specifications on
Bars and Extruded Section of Aluminum-Zinc-Magnesium-Copper-Chromium
Alloys. Ministry of Defence, UK, 1973.
IS 4904: Calibration Blocks for Use in Ultrasonic
Nondestructive Testing - Specification. Bureau of Indian Standards, 1990.
- * Ultrasonic Section, National Physical Laboratory, Dr. K.S. Krishnan
Road, New Delhi 110 012, India; 91-11-5787161; fax 91-11-5752678;
e-mail npl@sirnetd.ernet.in.
Copyright © 1997 by the American
Society for Nondestructive Testing, Inc. All rights reserved.
[ Materials Evaluation
]