|
The measurement of any parameter is of no
value without some certainty that the measurement is of sufficient
accuracy. This NDT Solution paper provides useful insight to
ultrasonic calibration. Understanding of the process can ensure
that you can specify and obtain meaningful calibration and,
ultimately, reliable ultrasonic measurements.
G.P. Singh
Contributing Editor
|
Background
The
normal practice in most test methods is the calibration of the equipment
to ensure the accuracy and repeatability of the test. Industry standards
are usually invoked to give guidance regarding the calibration and adjustment
of testing equipment. The document MIL STD 45662A: Military Standard
for Calibration System Requirements, has stood alone for many years
as the means by which a calibration system should be set up and applied.
New standards have now emerged to replace MIL STD 45662A. Two such documents
are International Standard ISO 10012-1: Quality Assurance Requirements
for Measuring Equipment (metrological confirmation system for measuring
equipment), and American National Standard ANSI/NCSL Z540-1: Calibration
Laboratories and Measuring and Test Equipment - General Requirements.
These documents provide, to some degree, guidance
for selecting intervals for the frequency of calibrations or instrument
checks, personnel, traceability, reference materials, environment, procedures,
and records. These documents, however, do not give any direction on
how to calibrate specific equipment. They simply state that some interval
be selected and that some standard for calibration be established based
on usage and stability, with a specified measurement uncertainty and
accuracy. To specify to a calibration laboratory that an ultrasonic
instrument, for example, be calibrated in accordance with the requirements
of MIL STD 45662A is totally missing the intent of the document. Any
calibration laboratory that provides calibration based on this requirement
does not understand the calibration requirements or they may simply
invoke what they believe to be normal calibration limits (at times this
is stated as manufacturer's standards) or their own values that they
have established.
This area is somewhat ambiguous for the normal quality control
person in charge of calibration.
The actual guidance for calibration is normally
stated in the controlling specification or standard which specifies
ultrasonic inspection. The controlling document usually indicates the
calibration tolerances or limits as well as the calibration frequency.
Typical examples of these controlling documents are MIL STD 2154, AWS
D1.1, ASME Section V, Articles 4 and 5, as well as several ASTM test
methods, standards, or practices. These standards specify the extent
to which calibration is required. They may require only vertical and
horizontal linearity, or they could be quite extensive and require numerous
tests such as amplitude (dB) linearity and dynamic range, near and far
surface resolution, vertical and horizontal limits as well as linearity,
signal to noise ratio, and sensitivity.
The extent to which calibration is taken seriously,
or is even understood, depends upon the experience and dependability
of the person or company performing the checks or tests on the instrument.
The answers to the following questions can affect the reliability of
the calibration.
- Does the user have the proper calibration equipment
and/or test blocks?
- Is ultrasonic testing a regular requirement or
is it performed infrequently?
- Does the user have an established calibration
program?
- Does the user understand why the calibrations
are required?
- Does the user know what calibrations and methods
to specify, if performed by an outside source?
- Does the calibration source or facility, if an
outside organization, follow the customer's requirements?
- Does the outside organization know the manufacturer's
specifications or have the equipment manuals
- Does the outside organization have NIST traceability?
Calibration Mystery
One area that causes confusion is the definition of "acceptable"
calibration. If an ultrasonic instrument is sent out for calibration
for conformance to ASTM E317 and an electronic evaluation in accordance
with ASTM E1324 is made, does this mean that the instrument is calibrated
and acceptable for the work to be performed? Or, what if the instrument
is sent out with the statement "calibrate ultrasonic unit"
and an electronic or manufacturer's calibration is performed? This is
an area that is somewhat ambiguous for the normal quality control person
in charge of calibration, or even the ultrasonic technician who may
have to evaluate the results of the calibration or checks made on the
instrument by an outside calibration laboratory.
Figure 1 - Basic configuration for a resolution
block, from ASTM E317. Reprinted with premission of ASTM.
Basically, two different methods exist for calibration.
The first method is the use of electronic calibration or manufacturer's
specification to ensure that the ultrasonic unit is operating within
the specifications developed by the instrument manufacturer. The second
method is the evaluation of the performance characteristics of the entire
ultrasonic system. The two methods of calibration are actually designed
to work together as complements of one another. The scope of ASTM E1324
describes the intent of the electronic calibration and indicates that
it is to be used in conjunction with ASTM E317.
Electronic Calibration
The electronic calibration is used for several different reasons. One
reason is to verify that the instrument is working properly within the
manufacturerÕs specifications for each electrical circuit in
the system. These are measured against an electrical load or signal
by the use of an oscilloscope, multimeter, or other electronic device.
Calibrated external attenuators are also used to drive the applicable
sections of the circuits and provide specific input. In any case, the
limits or range of acceptable values must be specified.
The second reason an electronic calibration might
be used is to compare the response between two different ultrasonic
instruments. This could be required when attempting to duplicate test
results or when comparing instruments to determine which unit would
produce the best response in a specific test situation.
Areas that are examined electronically are gain,
linearity, pulse amplitude, time of flight peak mode, gate triggering,
noise level, cross talk, pulser timing, pulse power, pulse duration,
damping switch and filtering, time corrected gain, voltage input and
output, as well as a host of other characteristics.
System Calibration
The system calibration, in accordance with ASTM E317, is used to verify
the performance characteristics of the ultrasonic unit as used in conjunction
with a coupling medium, transducers, coaxial cables and connectors,
and test pieces or test blocks that are acoustically similar (heat treat
condition, grain size, grain flow, surface finish) to the material to
be tested. The scope of ASTM E317 also indicates that performance limits
must be specified. These are found in the specification requiring ultrasonic
inspection.
The calibration checks of vertical and horizontal
limits and linearity are fairly straightforward in application as well
as the accuracy of gain controls (Birks et al., 1991). The near and
far surface resolution, however, require that the customer specification
give direction as to what the minimum limits are.
No standard test blocks exist for every condition
that may be specified, so special blocks may need to be manufactured
to verify the resolution capabilities of the system. For instance, MIL
STD 2154 requires that the entry surface resolution is 3.175 mm (0.125
in.) for materials other than forgings for thicknesses up to 31.75 mm
(1.25 in.). This requires a 0.4 mm (0.015 in.) flat bottom hole at a
3.175 mm (0.125 in.) metal path. Many aerospace parts manufactured today
are designed with a minimum machining envelope to save money on expensive
materials such as certain stainless steels, titanium, Inconels, Hastelloys,
and other exotic materials. At times the customer requirement may be
as little as 1.3 or 1.5 mm (0.05 or 0.06 in.) near surface resolution.
The use of test blocks specified in MIL STD 2154 for calibration could
completely miss the requirement for near surface resolution. If an outside
source is used and only an electronic calibration is performed, the
requirement could not be verified. In addition, if external calibration
is performed, the specific limits would need to be delineated with regard
to surface resolution requirements.
Many ultrasonic units used for field testing do
not have the response necessary to distinguish near or far surface resolution
in some materials used in industry. Many high temperature materials
are highly attenuative, and noise levels could be too high for many
instruments to resolve. Figure 1, from ASTM E317, provides the basic
configuration for a resolution block. The hole diameters, metal travel
distances, and surface finish requirements must be specified by the
user.
Calibration Verification
When calibration is performed by an outside source, the equipment used
for calibration should itself be calibrated at specific intervals and
should be traceable to the National Institute of Standards and Technology,
or NIST (formerly the National Bureau of Standards, or NBS) if a standard
exists. Otherwise, the traceability should be to a recognized industry
standard.
Proper evaluation of a calibration source requires
more than a statement that indicates the instrument was calibrated to
NIST. The NIST traceability numbers should be listed or should be available
for each apparatus used in calibration. The NIST number should also
have a date indicating the actual date the instrument was compared against
the NIST standard. A master instrument or standard is typically recalibrated
every one to five years, based on the type of instrument, its stability
and the amount of use it receives, and the original equipment manufacturer's
recommendations. Care should be exercised in this review, since it has
been found that some sources have used NBS traceability numbers for
up to twenty years without performing recalibration. In any case, the
calibration source should have a list of the frequencies at which masters
are sent for comparison to NIST standards.
Another situation that can lead to inaccuracies
is when a calibration source uses a NIST number and its instrument may
not have been directly compared to a NIST standard. In other words,
the standard to which the calibration source compared its equipment
may have been compared to another instrument that was compared to a
NIST standard. This "chain" between the master instrument
and the NIST standard can become quite long at times, with three or
four "links." This tends to allow for inaccuracies in the
calibration, since the deviations may build up. When a number of links
are used, each link should identify the amount of uncertainty that is
inherent in the calibration process. Typically, a five to one ratio
is used: the outside organization must have its equipment calibrated
to an accuracy of two percent in order to provide an accuracy of within
ten percent to its customers.
Conclusion
The calibration of ultrasonic instruments and the standards used for
the calibration both require a detailed review of the applicable specifications
invoked by customers. An understanding of the different means of calibration
is important because the use of the wrong calibration requirements could
invalidate many hours of inspection work. Electronic calibration should
be used in conjunction with a system performance evaluation, since the
two methods complement each other. One logical means to accomplish calibration
is to conduct the performance based calibrations on a set interval and
augment these with the electronic calibration on a semi-annual, annual,
or other basis as determined by equipment use, customer requirements,
and stability. In any case, the performance based calibration should
take precedence.
Many of the requirements for ultrasonic equipment
calibration were based on the response available from older instruments.
The values listed in many documents are no longer valid since newer
units can easily pass the requirements. Industry on a large scale is
moving toward performance based standards. ISO documents are rapidly
being developed and are taking the place of outdated standards that
have been around for many years. Input from equipment manufacturers,
companies specifying ultrasonics, and users is needed to provide industry
with viable standards that will be practical for use with the newer
equipment now available. Some manufacturers have already introduced
computer controlled calibration checks that electronically drive the
instrument and provide a report with the calibration values and limits
listed. This is a giant stride toward simplifying and standardizing
the calibration checks. A similar consolidation of methods and requirements
is needed in the performance based calibration area. Of course, each
industry has particular needs, and some may not want strict limits that
could increase the costs of calibrations without providing significant
benefit to the test. The education of technicians in the area of equipment
calibration could also help to provide calibration laboratories with
the correct information and proper direction needed for satisfactory
calibration.
References
ASTM E1324: Standard Guide for Measuring Some Electronic Characteristics
of Ultrasonic Examination Instruments.
ASTM E317: Standard Practice for Evaluating Performance
Characteristics of Ultrasonic Pulse-Echo Testing Systems.
Birks, Al, Robert E. Green, Jr., and Paul McIntire,
Nondestructive Testing Handbook, Volume 7, Ultrasonic Testing,
2nd ed., 1991, pp 465-470. ASNT, Columbus, OH.
| * |
Alloyweld Inspection Company, 796 Maple Ln., Bensenville,
IL 60106; (630) 595-2145 |
| + |
TPH Enterprises, Inc., 334 E. First St., Genoa, IL 60135;
(815) 784-4784. |
| # |
AlliedSignal Engines, AlliedSignal, Inc., 400 S. Buncombe
Rd., Greer, SC 29651; (864) 989-3881. |
Copyright © 1996
by the American Society for Nondestructive Testing, Inc. All rights
reserved.
[ Materials Evaluation
]