
| Volume 2, Number 3 |
|
July 2003 |
TNT
FYI:
Reference and Calibration Standards
To ensure accurate and repeatable inspection,
ultrasonic testing equipment must be standardized and calibrated so that
data taken by different operators are comparable and can be matched
against inspection norms. This is accomplished through the use of
reference and calibration standards or test blocks.
Selecting a Reference Standard. Selection
of a reference standard is determined by testing technique, type of
material to be inspected and form, type of discontinuity to be detected,
and specification requirements. Reference standards, such as those
designed to ASTM, ASME or AWS specifications, are used to standardize
equipment responses. Blocks with flat bottom holes (Fig. 1) are often
used to standardize amplitude of the detected signal with respect to the
effective area or distance of known reflectors. Area (amplitude or
distance) amplitude curves are usually constructed using such blocks. In
some cases, blocks with side drilled holes are also used for such
standardizations. Reference standards varying acoustically from the test
object by more than 6 decibels are usually considered unacceptable for
use.
Preparing a Reference Standard. Sometimes,
it is preferable or required to prepare a reference standard from a
piece of the same material as that to be tested, by introducing notches
or holes into a sample or into the actual test object. The advantage of
such a reference standard is that the test object and the standard will
have the same composition, manufacturing history, surface condition and
geometry. the disadvantages are that usually there will be fewer
artificial reflectors and it may not be possible to manufacture the
reflectors as accurately as might be done with a separate standard.
Other types of standard test blocks (and some of
the above reference blocks) are used to calibrate the ultrasonic testing
equipment with respect to essential variables such as sweep length,
pulse energy and amplification, search unit characteristics,
sensitivity, resolution and linearity. Typical calibration blocks
include IIW type blocks, DC, SC, DSC and MAB blocks. Two uses of the IIW
type blocks are shown in Fig. 2a and b.
Conclusion. In all cases, reference
standards must be prepared and used in strict accordance with well
designed specifications that cover the material, the fabrication and the
application of the blocks.
Figure 1. Reference
block diagrams (a) and (b) illustrate flat bottom hole for are
and distance amplitude blocks, (c) side drilled, (d) known dimension
notch and (e) flat bottom hole in blocked manufactured from actual
part. |
|
Figure 2. Two uses
of the IIW type block include (a) IIW verification of angle beam
search unit beam index point and (b) IIW determination of straight
beam dept resolution |
|
[ The NDT Technician ]
Copyright © 2008 by the American Society for Nondestructive Testing, Inc. ASNT is not responsible for the authenticity or accuracy of information herein. Published opinions and statements do not necessarily reflect the opinion of ASNT. Products or services that are advertised or mentioned do not carry the endorsement or recommendation of ASNT.
IRRSP, NDT Handbook, The NDT Technician and www.asnt.org are trademarks of the American Society for Nondestructive Testing, Inc. ACCP, ASNT, Level III Study Guide, Materials Evaluation, Nondestructive Testing Handbook, Research in Nondestructive Evaluation and RNDE are registered trademarks of the American Society for Nondestructive Testing, Inc.
ASNT exists to create a safer world by promoting the profession and technologies
of nondestructive testing.